3.6.1.
Port Config Menu-Wide Controls
The
Read Settings
button is the only
menu-wide operation in the
Port Config
menu.
Read Settings
is used to update
all
Port Config
menu configuration
settings to match actual resource settings
in the current test port of the currently
connected instrument. Generally, this
operation is not frequently necessary
because each time a new test port is
selected in the Slot-Port helper menu or
each time the
Port Config
tab menu is
selected, all displayed settings will
update automatically to match actual
resource settings in the test port.
The
Read Settings
button will also clear
out any pending configuration changes
and assure synchronization of the Port
Config tab menu to associated test port
resources.
3.6.2.
Port Config Actions Sub-Menu
The
Actions
sub-menu within the
Port Config
tab menu (
see Figure 3.10
) is used to:
1.
Apply Port Connection configuration changes
2.
Apply Port Impairments configuration changes
3.
Copy Port Connection, Port Impairment, and/or Active Impairment configuration settings to other test ports from
the current test port
Whenever radio buttons are selected in the
Port Connection
and
Port Impairments
frames of the Port Config menu, the
Apply Selected Port Settings
button will
annunciate
red
to indicate that a pending change should be applied (
see Figure 3.11
).
At this point, pressing the
Apply Selected Port Settings
button will effect the
selected configuration setting in the current test port if the corresponding checkbox for
Port Connections
and Port Impairments
is also selected.
The three checkboxes,
Port Connections
,
Port Impairments
, and
Active
Impairments
serve as filters for which settings can be applied or replicated to other
test ports. For example, if
Port Connections
is de-selected, then any settings selected
under the Port Connection frame will not be applied when the
Apply Selected Port
Settings
button is pressed and will not be copied when the
Copy Selected Settings
button is pressed.
Replications of configuration settings will be
further discussed below in Section 3.7.
3.6.3.
Port Connection Sub-Menu
The
Port Connection
sub-menu governs the switch position of the PVA Switch
in Figure 2.1.
Test PHY
causes the
TEST
interface to connect to the TEST
PHY while
THRU Port
routes the
TEST
interface via the 4-pair coupler to the
THRU
interface. When either of these radio buttons are selected, the
Apply
Selected Port Settings
button in the
Actions
frame will annunciate
red
(
see
Figure 3.11
) meaning that the configuration change will only take effect when
that button is pressed
and
if the
Port Connections
checkbox is selected.
3.6.4.
Port Impairments Sub-Menu
The
Port Impairments
sub-menu also controls test port switching options. The
100 Meter Cat5e
sub-menu represents the
Line Loss
impairment discussed
Figure 3.10
Actions Sub-Menu
Figure 3.11
Pending Port Config
Figure 3.9
Port Configurations Tab Menu
Summary of Contents for PhyView PVA-3000
Page 22: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 22 ...
Page 70: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 70 ...
Page 108: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 108 ...
Page 124: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 124 ...