Sierra Instruction Manual
InnovaSonic
®
203 Economical
12
You can check/enter numerical data by visiting the Data Type menus. If you want to modify the data, input the
digits and press
or press
first, input the digits then press
again to confirm.
Example 1: To enter a pipe outside diameter of
219.234, the procedure is as follows:
Press
keys to enter menu M11 (the
numerical value displayed currently is the previous
value). Now press
key. The symbol “>” and the
flashing cursor are displayed at the left end of the
second line on the screen. Then input pipe OD and
press enter to confirm.
.
You can select various configurations by using pull down menus in the Option Type menus. If you want to modify
these, press
first, the symbol “>” and the flashing cursor are displayed at the left of the screen. Use
and
to scroll pull down menu options, then press
to confirm the selected option, or enter the
corresponding value directly and press
to confirm.
For example, if the pipe material is not “Stainless
Steel”, press
to enter menu M14, press
to modify the options. Scroll to “1. Stainless
Steel” option by pressing
and
, then press
to confirm the selection; It is also possible to
press
key to change the selection and wait until
“1.Stainless Steel” is displayed on the second line of
the screen, then press
to confirm.
1.6.
Flow Meter Menu Descriptions
Menus are assigned as follows:
00~09
Flow Totalizer Display: display flow rate, positive total, negative total, net total, velocity, date & time,
present operation and flow results today.
10~29
Initial Parameter Setup: to enter pipe outside diameter, pipe wall thickness, fluid type, transducer type,
transducer mounting method and spacing, etc.
30~38
Flow Units Options: to select the flow unit such as cubic meters, liters or other units, turn totalizers
on/off and reset totalizers, etc.
40~49
Setup options: set Scaling factor, system lock (Menu M47), etc.
50~89
Input and output setup: date and time, ESN, communication baud rate setting, etc.
90~94
Diagnostics: Signal strength and signal quality (Menu M90), TOM/TOS*100 (Menu M91), flow sound
velocity (Menu M92), total time and delta time (Menu M93), Reynolds number and factor (Menu
M94), etc.
Pi p e Ou t e r Di a me t e r
>_
ATTENTION
Generally, press the
key first if you want to change data. If there is no response “, it means that
system is locked by a password. To “Unlock” it, select “Unlock” in Menu M47 and enter the original password.
Pi p e Ma t e r i a l [ 14
>1. St a i n le s s St e e l