Connecting
4.2 Connecting note
Transducers
Compact Operating Instructions, 09/2021, A5E43390688-AC
33
NOTICE
Condensation in the device
Damage to device through formation of condensation if the temperature difference
between transportation or storage and the mounting location exceeds 20 °C (36 °F).
•
Before taking the device into operation let the device adapt for several hours in the new
environment.
Note
Electromagnetic compatibility (EMC)
You can use this device in industrial environments, households and small businesses.
Note
Improvement of interference immunity
•
Lay signal cables separate from cables with voltages > 60 V.
•
Use cables with twisted wires.
•
Keep device and cables at a distance from strong electromagnetic fields.
•
Take account of the conditions for communication specified in the Technical specifications
(Page 38).
•
Use shielded cables to guarantee the full specification according to
HART/PA/FF/Modbus/EIA-485/Profibus DP.
4.2
Connecting note
Note
Installation should only be performed by qualified personnel and in accordance with local
governing regulations.