![Siemens SIREMOBIL Installation Instructions Manual Download Page 7](http://html.mh-extra.com/html/siemens/siremobil/siremobil_installation-instructions-manual_381866007.webp)
2 - 1
Siemens AG
SPR2-230.812.02
Page 1 of 8
SIREMOBIL
Medical Solutions
Rev. 02
02.06
CS PS SP
System Manual
Installation of TFT displays
2
Dismantling the existing monitors
2
•
Remove both monitors as follows.
•
Remove transport monitor covers (Fig. 1).
•
Remove the shielding plates (Fig. 2).
•
Remove the connection housing (only for SIMOMED N monitor 11 02 628, Fig. 3).
•
Unplug all cables from the monitor to be replaced.
•
In each case loosen the 2 screws at the monitor mounting (Fig. 4).
•
Lift down both monitors.
•
Reattach all previously removed parts to the monitor.
Fig. 1 Monitor cover
Fig. 2 Shielding plate
Fig. 3 Connection housing
Fig. 4 Monitor attachment
Mark for the keyboard tray
Mark for the keyboard tray