![Siemens SIREMOBIL Installation Instructions Manual Download Page 15](http://html.mh-extra.com/html/siemens/siremobil/siremobil_installation-instructions-manual_381866015.webp)
3 - 1
Siemens AG
SPR2-230.812.02
Page 1 of 2
SIREMOBIL
Medical Solutions
Rev. 02
02.06
CS PS SP
System Manual
Settings
3
TFT display BARCO MVG 1318, part number 80 79 530
3
•
Make the checks and if necessary settings according to the replacement instructions
TD00-000.841.17...
Refer to the System-dependent settings chapter of the instructions.
Concluding work
3
•
Close all covers again. Make sure that the protective ground wire connections make
good contact.
•
Perform the protective ground wire test according to TD00-000.860.01.
The protective ground wire resistance must not exceed 0.2 ohms.
•
Perform the IQ quick test of the system.
•
Perform the country-specific acceptance tests (part acceptance).