28
ABLB
Alternate Binaural Loudness Balance: Routes
tone to each earphone alternately. Use the SI-
GNAL rotary control to adjust the Left channel
level and the MASK rotary control to adjust the
right. Pressing the PRESENT key interrupts the
signal presented.
SISI
Short Increment Sensitivity Index: Presents a
tone continuously at the level and frequency dis-
played. Pressing the PRESENT key increases the
level by the increment shown on the lower line
of the display. Change this increment by using
the UP & DOWN softkeys. Softkey 4 switches bet-
ween the initial display and a counter that logs
and displays the results of the test as a percen-
tage: