background image

Multi-Echo EPI 

Page 8 of 9 

Numaris/4 VA21B 

 

Separate Correction

: In the source code of the sequence card a new parameter 

‘Separate Correction’ has been introduced. If this parameter is enabled (during 
compilation) each echo image is preceded by a new set of navigators used for the 
Nyquist ghost correction by the reconstruction. If the feature is disabled the same 
correction is used for all echo-images. The echo times are somewhat shorter when 
this feature is disabled, however, in general the image quality is better with the 
separate correction enabled (especially for high imaging bandwidth). In rare cases 
the signal in the later echoes is too low and the reconstruction introduces artifacts in 
the images (small noisy stripes in the image) if this feature is selected. Please contact 
the developer if changes are needed. 
 

T2*-Fit (parameter not shown)

: This new feature allows online calculation of I0- and 

T2*-maps from the echo images according to the model function shown in 1.2. I0 and 
T2* are determined for each voxel who’s intensity exceeds 8% of the maximum 
intensity in the first echo. I0 and T2* of voxels with a signal intensity below this 
threshold are set to 0. 

This option is automatically and only enabled when 

motion correction is selected

. The resulting T2*-maps are scaled such that an 

image intensity of 1000 corresponds to a T2* of 100ms. T2* is forced to lie within 
physiological limits (0<T2*<333ms). Values higher than 333ms are set to 334ms to 
avoid very large signal intensity differences in case of fitting errors and allow a T2*-
resolution of 0.1ms without overflows. Therefore, T2* maps of phantoms with very 
long T2 and good homogeneity may be calculated falsely. 

The T2*-fit is only 

performed when Mosaic-images are created (for more than 1 slice)!

 

 

Motion Correction

: Similar to the motion correction in the a_ep2d_mosaic sequence 

a full 3D motion correction can be performed online. In the Multi-Echo EPI sequence 
this option also enables the T2*-Fit since the parameter maps are needed for motion 
correction. For most accurate results the motion correction is based on the I0-maps, 
which are used to determine the motion parameters. Subsequently these parameters 
are applied to reorient the T2*-maps. Only the motion corrected T2*-maps are written 
to the database together with the original images. 
The 

PACE

-variant of the sequence (separate sequence: os_mepi2d_pace_VA21) 

allows prospective motion correction. Thus, the volume is rotated and shifted after 
each measurement to ensure acquisition in the same patient based reference frame. 
Residual motion is corrected for in the processing. This option is only enabled when 
the motion correction is selected, which in turn enables the T2*-fit (which is 
necessary for motion correction (see above). Be aware that the prospective motion 
correction stops if the motion between two scans is too large (as in the product EPI-
PACE sequence). 
 

Interpolation

: The interpolation should be set to ‘k-Space’! (see 3D Filter) 

 

3D Filter

: The filter has the same function as in standard EPI (remember: smaller 

values cause stronger spatial filtering!). 
 

Summary of Contents for MULTI-ECHO EPI 4A21

Page 1: ...etom under VA21B Installation and User s Guide NUMARIS 4A21 April 4th 2003 Section of Medical Physics University Hospital Freiburg Germany Contact Oliver Speck PhD Hugstetterstr 55 D 79106 Freiburg email oliver speck uniklinik freiburg de phone 49 761 2703832 ...

Page 2: ...ciple of Multi Echo EPI 3 1 2 Image contrast of Multi Echo EPI 4 1 3 Processing options 4 1 4 References 4 2 Software Installation Procedure 5 2 1 Installation 5 3 Sequences and Protocols 7 3 1 General Remarks 7 3 2 New Parameters 7 3 3 Remarks on reconstruction performance 9 3 4 Protocols 9 ...

Page 3: ...se gradient is rewound to the original starting position after each echo image to ensure identical k space trajectories for all echo images 2 GP RF GS GR echo image loop n Figure 1 Basic timing of the Multi Echo EPI sequence The shaded gradient in phase encoding direction ensures identical k space trajectories for all echo images The resulting echo images have different increasing echo times and t...

Page 4: ...ifferent echo times 3 Please see the processing options described below Figure 2 Head images acquired on Magnetom TRIO with the Multi Echo EPI sequence with 4 different echo times in a single shot 1 3 Processing options Several processing options are given in the method The different echo images are always saved separately as the original images and optionally parametric maps can be calculated in ...

Page 5: ...mepi2d_VA21 m750 n4 ppc750 prod bin Attention This is only correct for systems with Reco PC and m750 MPCU Otherwise copy the relevant MPCU file to the corresponding directory Software does not run on Alpha reconstruction machines CoilSelectManipulator h ist ein noch fehlendes File in C idea n4 comp Measurement CCSpuSer CoilInterface Copy the EVA program dll files to the customer EVA directory copy...

Page 6: ...Multi Echo EPI Page 6 of 9 Numaris 4 VA21B ...

Page 7: ...tor will sometimes not allow the measurement to be performed with the parameters selected In general a lower bandwidth will decrease the stimulation and a different selection of the readout gradient may also help read direction along L R axis shows generally the lowest stimulation 3 2 New Parameters Number of echoes The number of echo images per excitation can be set on the sequence card part1 as ...

Page 8: ... set to 334ms to avoid very large signal intensity differences in case of fitting errors and allow a T2 resolution of 0 1ms without overflows Therefore T2 maps of phantoms with very long T2 and good homogeneity may be calculated falsely The T2 fit is only performed when Mosaic images are created for more than 1 slice Motion Correction Similar to the motion correction in the a_ep2d_mosaic sequence ...

Page 9: ...re long reconstruction times after the scan are possible The scan is not aborted However no new scan can be started during reconstruction Do not move the patient table to the home position during reconstruction otherwise the reconstruction will stop and data is lost The number of images that can be processed online depends on the reconstruction hardware Up to 20 slices with 4 echoes and a TR of 3s...

Reviews: