Recalibration
7.4 Recalibration
BW500 and BW500/L
82
Operating Instructions, 12/2016, A5E33482052-AD
Connections
If the span selection is to be done by remote contact, the following connections would apply.
Otherwise, no additional connections to the BW500 are required.
Multispan Selection of Spans 1 and 2
Multispan Selection of Spans 1 to 8
*Remote contact can be from relay, open collector switch, or BCD switch.
7.4.8.1
Programming
Access P365 and enter EDIT mode
P365 Multispan
Select [1-8]
E
0
Span 1 will have already been set as part of the Start Up and initial calibration. Therefore, select
2.
Access P017 and enter EDIT mode
P017 Test Load: Weight MS2
Enter test load
E
0
Enter the test load value, and press the following key to do a span calibration:
To do a span calibration for another condition, (i.e. span 3 or 4 etc.), access P365 and repeat
these steps for each condition. As with any initial span, follow the span calibration for each
multispan with a material test and factoring.
To use remote span selection, auxilliary Inputs, 1 and/or 2 or 3, are programmed to read the
contact state as the span selection. Remote selection overrides the keypad (or Dolphin Plus)
selection. The auxilliary inputs override the keypad selection.
Summary of Contents for Milltronics BW500
Page 1: ...Operating Instructions Milltronics BW500 and BW500 L 12 2016 Edition Integrators ...
Page 12: ...Table of contents BW500 and BW500 L 10 Operating Instructions 12 2016 A5E33482052 AD ...
Page 14: ...Safety Notes BW500 and BW500 L 12 Operating Instructions 12 2016 A5E33482052 AD ...
Page 16: ...The Manual BW500 and BW500 L 14 Operating Instructions 12 2016 A5E33482052 AD ...
Page 48: ......
Page 154: ......
Page 218: ...Index BW500 and BW500 L 216 Operating Instructions 12 2016 A5E33482052 AD ...
Page 219: ......