8 Miscellaneous
SIE
Testing the radiation field / light field
•
Proceed with the test according to the Installation & Start-Up instructions
SPB7-115.814.01… chapter "Checking the radiation field limitation".
PMF Automatic filter selection
•
If it exists select it from the control panel and check that audible sound during movement is
normal.
PMF RX signal lamp
•
Does the radiation-ON indicator light up?
PMF AEC Detector test
•
Check the Sensibility test [r] parameter as following:
During the test, the system should be configured as shown in the figure below.
•
Install Grid table 18x24
•
Install 18x24 collimation plate
•
Select LARGE focus
•
Remove compression plate
•
Remove cassette
•
Select ZERO POINTS technique