5 Rotating anode starter
MAMMOMAT Balance
SPB7-115.831.01 Page 13 of 22
Siemens AG
Maintenance Instructions
Rev.01 05.04
Medical Solutions
PMF Control LED (brake disabled)
•
PCB 92-074
LED1
(yellow) bright during normal rotation phase (RUN)
LED2
(yellow) dark during breaking or alarm
LED3
(yellow) dark during breaking
LED5
(green) bright in stand by or rotation
LED6
(green) bright during rotation
PMF Brake disabling for heavy use
•
If upon log file inspection high duty comes in evidence it’ recommended to disable anode rotation
brake to reduce operative tube assembly temperature.
For more details refer to service instructions SPB7-115.840.01….