7 High tension power supply
MAMMOMAT Balance
SPB7-115.831.01 Page 17 of 22
Siemens AG
Maintenance Instructions
Rev.01 05.04
Medical Solutions
PMF Checking and adjustment points
TP1-TP2 for kV checks on (PCB 94-132)of inverter block 08883436
RV1 for kV correction
PMF CONTROL LED
Green LED1 indicates that fuse F2 of PCB 94-132 is OK.
Yellow LED 2 turns on during exposure and can be used to verify continuous or pulsated X-ray
emission.