46
Siemens-Elema AB
E382 EH81E 061 01 03 02
Troubleshooting
Compressor Compact
Malfunction
Action
Water appears to be leaking from the
Compressor Compact:
• Inspect all water tubes and connectors.
Tighten loose connectors and/or replace
damaged components.
• Inspect the drainage bottle for leaks or a
loose connection. Replace the bottle if
necessary.
• Inspect filter/water separators and replace
them if necessary.
• Inspect solenoid valves and replace them if
necessary.
Water does not accumulate in the drainage
bottle at regular intervals. Drainage is too
frequent, not frequent enough, or does not
occur at all:
• Inspect all water tubes and connectors.
Clear obstructions and/or replace faulty
components.
• Inspect filter/solenoid valves and replace
them if necessary.
• Inspect water separator 2 and replace it if
necessary.
• Replace the control section PC board.
• Malfunction in the thermoelectric cooler.
Check the function of the cooler temp.
switch as previously described. Replace
the thermoelectric cooler if necessary.
The operating time meter does not run when
the Compressor Compact is in operation:
• Replace the operating time meter.