34
Siemens-Elema AB
E382 EH81E 061 01 03 02
Disassembling and assembling
Compressor Compact
Filter/solenoid valves
To remove the filter/solenoid valves
• Open the front cover and remove the rear
cover and the casing as described in the
section Compressor covers.
• Loosen the screw (1) and slide out the
filter/solenoid valve unit (2).
• Hold the filter/solenoid valve unit while
unscrewing the couplings (3). Use a
21 mm open-end wrench.
Note
– The restriction washers (4), the
mesh nets (5) and the plastic rings (6)
must be replaced at the 8000 hour
overhaul.
To refit the filter/solenoid valves
• The procedure for refitting the filter/
solenoid valves is the reverse of the
removal procedure previously described.
2
1
COMP019X
4
5
6
3
COMP028X
6