Description of functions
Compressor Compact
3.12 Fuses
The compressor is equipped with three
FUSES: two mains power fuses (F1 and F2)
and one fuse for the 12 V power supply (F3).
The fuses are located in the 3.11 ELECTRI-
CAL INTERCONNECTION block.
Fuse No. Type No.
Manufacturer Dimension
115 V versions
Mains power fuses:
F1/F2
MDA 10A 250V Bussman
6.3 x 32
12 V power fuse:
F3
MDA 15A 250V Bussman
6.3 x 32
230 V versions
Mains power fuses:
F1/F2
T 6.3AL
–
5 x 20
12 V power fuse:
F3
T 10AL 250V
Littelfuse
5 x 20
When replacing fuses in the Compressor
Compact, make sure to use fuses as stated
above (Type No. and Manufacturer).
If the compressor is connected to mains
power, there is live voltage at the mains
power fuses even when the on/off switch
is in OFF (0) position.
3.13 Dust filters
There are two DUST FILTERS mounted on
the compressor housing. These two identical
filters are made of a synthetic material.
They must be regularly inspected and
cleaned as described in the Operating
Manual, chapter Cleaning. The filters must
be replaced at the 2000 hour overhaul as
well as at the 8000 hour overhaul.
3.14 Fan
The Compressor Compact is equipped with
a double inlet centrifugal FAN. The purpose
of the fan is to provide forced air cooling for
the 1. COMPRESSOR SECTION and the
2. COOLING SECTION.
There are two different versions of the fan,
one for 115 V mains power supply and one
for 230 V mains power supply.
The fan motor is fitted with a self-resetting
thermal switch that will cut the power supply
to the fan motor if the temperature at the
switch exceeds approx. 130°C (266°F).
COMP045X
F1 / F2
F3