Siemens 66 Series Installation And Service Instruction Download Page 4

SD66 

 

4

MAINTENANCE 

 
Most problems associated with pneumatic instruments can be prevented by providing clean, dry, oil-free instrument 
air (see Instrument Air Requirements). If these requirements are met, routine maintenance should not be necessary. 
Cleaning of the plunger and the supply and exhaust ports may be required, on an occasional basis, as outlined in the 
Cleaning section. 
 

CAUTION

 

Before disassembling the instrument, turn off supply air pressure to the instrument. 

Troubleshooting 

 
Table 1 provides troubleshooting information in a symptom, cause, and remedy format. Symptoms are listed in the 
order they are most likely

 

to

 

occur. 

 

Table 1  Troubleshooting 

 

SYMPTOM CAUSE 

REMEDY 

Plunger is not seating properly in 
supply port due to dirt, chips, etc. 

Refer to Maintenance, Cleaning and clean 
plunger and supply port. 

Exhaust port is clogged due to build up 
on the plunger and/or exhaust port. 

Refer to Maintenance, Cleaning and clean 
plunger and exhaust port. 

Output pressure will not 
decrease or decreases with a 
decrease in input pressure. 

Exhaust port is obstructed. Clear 

obstruction. 

Plunger is not seating properly in 
exhaust port due to dirt, chips, etc. 

Refer to Maintenance, Cleaning and clean 
plunger and exhaust port. 

Supply port is clogged due to build up 
on the plunger and/or supply port. 

Refer to Maintenance, Cleaning and clean 
plunger and supply port. 

Output pressure will not 
increase or increases with an 
increase in input pressure. 

Supply pressure is too low. 

Raise supply pressure. 

Ruptured input diaphragm or output 
diaphragm assembly. 

Refer to Maintenance to replace input 
diaphragm or output diaphragm assembly. 

Excessive exhaust from 
exhaust vent. 

Plunger is not seating properly in 
exhaust port due to dirt, chips, etc. 

Refer to Maintenance, Cleaning and clean 
plunger and exhaust port. 

 

Cleaning 

 
To clean the plunger and the supply and exhaust ports: 

1.

 

Remove supply and input pressures. 

2.

 

Remove the retaining nut from the bottom forging (see Parts List). The plunger and plunger spring should fall 
out. 

3.

 

Clean the plunger with non-abrasive solvent. 

4.

 

Clean the supply and exhaust ports with non-abrasive solvent (see Figure 2). The supply port is readily 
accessible; the exhaust port can be reached with a tobacco pipe cleaner. 

 

NOTE 

If there is excessive buildup on the supply or exhaust port, the relay must be 
disassembled for thorough cleaning. Refer to the Parts List for parts orientation during 
disassembly and assembly. 

5.

 

Refer to the Parts List and install the plunger, plunger spring, and retaining nut. 

6.

 

Check the calibration of the Relay per the following Calibration. 

 

StockCheck.com

Downloaded from StockCheck.com

Summary of Contents for 66 Series

Page 1: ...eration Maintenance Warranty and Parts List Model Designation 66 BA 2 Model Series Material and Function A Brass Amplifying R Brass Reducing Input to Output Ratio 2 1 2 Model 66BA2 2 1 Model 66BR2 4 1 4 Model 66BA4 4 1 Model 66BR4 3 1 3 Model 66BA3 3 1 Model 66BR3 6 1 6 Model 66BA6 6 1 Model 66BR6 Specifications Supply Pressure Normal 20 psig Minimum 1 psig above maximum required output Maximum 80...

Page 2: ...may cause damage to the relay Pressures in excess of those stated in the Specifications section may cause damage to the relay Instrument Air Requirements Connect the instrument i e relay to a source of clean dry oil free instrument air Failure to do so will increase the possibility of a malfunction or a deviation from specified performance CAUTION Use of process fluids other than instrument air is...

Page 3: ...ive areas of the input diaphragm and output diaphragm assembly The input diaphragm of an Amplifying Relay has a larger effective area than the output diaphragm assembly which requires the output to be proportionally higher for the relay to be in balance see Figure 2 Amplifying View Conversely the input diaphragm of a Reducing Relay has a smaller effective area than the output diaphragm assembly wh...

Page 4: ... port Supply port is clogged due to build up on the plunger and or supply port Refer to Maintenance Cleaning and clean plunger and supply port Output pressure will not increase or increases with an increase in input pressure Supply pressure is too low Raise supply pressure Ruptured input diaphragm or output diaphragm assembly Refer to Maintenance to replace input diaphragm or output diaphragm asse...

Page 5: ...d plunger spring 7 Insert a 1 16 Allen wrench through the exhaust port in the bottom forging make adjustment 8 Install plunger plunger spring and retaining nut 9 Repeat steps 4 through 8 until span is set Function and Ratio Change The function and or ratio of a Relay can be changed using standard parts To change the function from amplifying to reducing or vice versa with the same ratio e g Model 6...

Page 6: ...formation is provided on the product nameplate part number or model number serial number and or version For most software this information is given in the Help About screen If there is a problem with product operation Is the problem intermittent or repeatable What symptoms have been observed What steps configuration changes loop modifications etc were performed before the problem occurred What sta...

Page 7: ...t the goods do not conform to the above warranty Repaired and replacement parts shall be warranted for the remainder of the original period of notification set forth above but in no event less than 12 months from repair or replacement At its expense Buyer shall remove and ship to Seller any such nonconforming items and shall reinstall the repaired or replaced parts Buyer shall grant Seller access ...

Page 8: ...is subject to change without notice Procedures in this document have been reviewed for compliance with applicable approval agency requirements and are considered sound practice Neither Siemens Energy Automation Inc nor these agencies are responsible for repairs made by the user IMPORTANT Service Parts Kits are available for servicing the instrument Contact Siemens for available kits refer to the P...

Reviews: