8.1 Daily checkup (by user)
This check could quickly verify if the equipment is suitable for the use and it takes few minutes.
1.
Connect the 6110-A1 and 6110-A2 to a main power supply.
2.
Switch on the 6110-A1 and 6110-A2, pushing the on/off switch (if the foot switch is present push the switch on position l),
the equipment must work smoothly without changing of motor turns, not normal noises or vibrations
3.
Close the VACUUM tube, turn in clock wise the Vacuum regulator and check that the VACUUM level reaches at least the value
indicated on following tables (less 5% max).
4.
With the VACUUM tube open, turn in clock wise the Vacuum regulator and check that the VACUMM level will decrease towards
0 (zero) (due to load loss could be accepted a residual vacuum level of 40-50 mbar).
5.
Switch off the unit.
6.
Check if the filter is clean and no liquid or other contamination inside. If the color of the filer is not white it must be replaced,
a dirty filter means not good performances of the 6110-A1 AND 6110-A2 (see the chapter 4.4).
7.
If necessary, replace filters and disposable bag (if used).
8.2 6/12 MONTHS CHECK UP (after warranty period)
This check verifies if the equipment is in compliance with the original productions features and so suitable to be used. This check
should be done by specialized operators or authorized service companies, Following the functional test this operators or companies
should do a safety electrical test in compliance with IEC 60601-1 and issue a final report.
TEST LIST TO BE DONE ON THE EQUIPMENTS:
1.
Check the pump: the Maximum Vacuum should be not less than the value indicated on following tables (less 5% max), and
there are not noises or vibrations.
2.
Check the vacuum regulator rotating it in both the senses and verifying smooth working operations.
3.
Check the mechanical integrity of the aluminium case, the penetration of external material could damage the equipment or
create danger to the operators.
4.
Check if the labels are still present and readable.
5.
Never open the device, for this technical operation please contact authorized service center only
6.
Check the Vacuum gauge. With the equipment off it must indicate 0 mbar (zero)
7.
Check the integrity of the canisters, connections and tubing.
8.
If necessary, replace filters and disposable bag (if used)
9.
Before declaring the equipment in compliance with the manufacturer features make a safety electrical test (IEC 60601-1). For
making this test ask to the authorized service center and/or to the manufacturer.
Use disposable or spare parts supplied from the manufacturer only, the compliance of disposable or spare
parts could be confirmed by the manufacturer only.
Issue and properly archive a suitable report for the check done.
The operator must consider that the use of the 6110-A1 and 6110-A2 in high quote, in this condition the
value of vacuum could reduced, do not use the 6110-A1 and 6110-A2 over 3000 mt from the sea level (9842 ft)
.
9 TECHNICAL SPECIFICATIONS
9.1 Transport/Storage conditions
The 6110-A1 AND 6110-A2 pump and accessories must remain in the packaging for storage and stored at a temperature
range from –25 °C to +70 °C (–13 °F to +158 °F).
9.2 Operating temperature
The 6110-A1 AND 6110-A2 pump and accessories must be operated within a temperature range of +5 °C and +40 °C
(+41 °F and +104 °F). Do not operate the products in extreme cold or heat.
9.3 Transport/Storage/operating conditions (humidity)
The 6110-A1 AND 6110-A2 pump and accessories must remain in the packaging for storage and used at a humidity
range from 15% to 93 %.
9.4 Transport/Storage/operating conditions (atmospheric pressure)
The 6110-A1 AND 6110-A2 pump and accessories must remain in the packaging for storage and used at an atmospheric
pressure range from 0,7 bar to 1,06 bar
CAUTIONS: Do not operate the 6110-A1 AND 6110-A2 pump above 3000 m above sea level (9,842 ft).
IP33
9.5 Protection class
The 6110-A1 AND 6110-A2 pump are protected against ingress of dripping water (IP33).