Siem Nova 6110-A1 CURET MAX User And Maintenance Manual Download Page 1

USER and MAINTENANCE MANUAL 

6110 A1 and 6110 A2 

Manufacturer: Siem Nova S.r.l. Legal and operational offices Via C. Varalli, 1 20087 ROZZANO (Milano) MADE IN ITALY 
Product: SURGICAL ASPIRATOR MOD: 

6110 A1 MAX - 6110 A2 SPEEDY                 

 

 

 

 

 

 

 

 

 

 

Updated: 2018-01-09 

 

reproduction forbidden 

Pag: 1 di 10 

 

 

 

 
ASSISTANCE 
 

Read the instruction manual before using the device.  
This manual must be kept with the device for later reference. 

 
 
 
 

SIGNS AND SYMBOLS  

 

 

This symbol indicates the 
compliance with the 
essential requirements of 
the Directive 93/42/ EEC 
of 14 June 1993 
concerning medical 
devices. Reference IMQ 
Spa - Milan      

 

This symbol 
indicates an 
insulation class I 
device.  

 

 

This symbol indicates 
to follow instructions 
for use. 

 

 

 

 

 

 

 

 

 

This symbol indicates the 
date of production (four 
digits for the year and two 
digits for the month). 

 

 

This Symbol 
indicates a type B 
applied part. 

 

 

This symbol indicates 
to consult instructions 
for use. 

 

 

 

 

 

 

 

 

 

This symbol indicates the 
manufacturer. 

 

 

This symbol 
indicates alternate 
current 

 

 

This symbol indicates 
the foot switch. 

 

 

 

 

 

 

 

 

 

This symbol indicates that 
interferences may occur in 
the vicinity of equipment 
marked with this symbol 

 

 

This symbol 
indicates a single 
use device. Do not 
reuse the device. 

 

 

This symbol indicates a 
CAUTION or WARNING 
associated with the 
device. 

 

 

 

 

 

 

 

 

This symbol indicates the 
humidity limitation for 
operation, transport and 
storage 

 

 

This symbol 
indicates the 
atmospheric 
pressure limitation 
for operation, 
transport and 
storage. 

 

 

This symbol indicates 
the temperature 
limitation for 
operation, transport 
and storage. 

 

 

 

 

 

 

 

 

 

This symbol indicates to 
keep the device away from 
sunlight 

 

 

This symbol 
indicates to keep the 
device dry. 

 

IP33 

This symbol indicates 
the protection against 
harmful effects due to 
ingress of solid foreign 
objects and against 
harmful effects due to 
the ingress of water. 

 

 

 

 

 

 

 

 

 
 

 

 

 

This symbol indicates do 
not dispose the device 
together with unsorted 
municipal waste (for EU 
only). 

 

 

 

 

 

This symbol indicates 
do not immerse in 
water or other liquid 

 

PRESSURE MEASURE UNIT CONVERSION TABLE 

 
Legenda:

 bar, KPa, cm Hg, cm H2O are all measure unit of pressure (vacuum)

 

 

bar 

KPa 

cm Hg 

cm H2O 

1 bar 

100 

75.006 

1019.72 

1 KPa 

0.01 

0.75006 

10.1972 

1 cm Hg 

0.133 

1.333 

13.595 

1 cm H2O 

0.098 

0.00098 

0.07355 

 
 

1

 

APPLICATION 

The  suction  equipments  6110-A1  and  6110-A2  are  approved  exclusively  for  the  use  as  described  in  this  user  and  maintenance 
manual. Siem Nova can only warranty the safe functioning of the equipments when the 6110-A1 and 6110-A2 are used with the 
Siem Nova accessories. 
 

      

Please read and observe these warning and safety instructions before use the device.  

       

Please note that these instructions for use are a general guide for the use of the product.  

 

Medical matters must be addressed by a physician.

 

 

Summary of Contents for 6110-A1 CURET MAX

Page 1: ...s symbol indicates the humidity limitation for operation transport and storage This symbol indicates the atmospheric pressure limitation for operation transport and storage This symbol indicates the t...

Page 2: ...er or liquids 24 Never touch with wet hands electrical parts 25 If the pumps get wet rub with dry towel Do not dry in a microwave 26 Do not touch the suction pumps when they have fallen into water Unp...

Page 3: ...ter 11x11 mm n 2 66609 Hydrophobic LIQUID STOP Filter n 2 66233 Silicon tube 7x13mm 4 m 81016 Pneumatic footswitch n 1 60777 Change over suction switch n 1 OPTIONAL 60972 Polycarbonate jar 2000 ml 900...

Page 4: ...FILTER 11X11mm ANTIVIRAL ANTIBACTERIAL HYDROPHOBIC FILTER SILICONE PIPE CONNECTOR YELLOW RUBBER CAP LIQUID STOP HYDROPHOBIC FILTER POLYCARBONATE CANISTER 2000ml PNEUMATIC FOOTSWITCH WHEELS WITH BRAKE...

Page 5: ...ging is damaged Do not reuse disposable use or sterile accessories 4 2 Assembly of the canister a check if the overflow valve is fixed on the lid of the canister and if the float with sealing is insid...

Page 6: ...ch on proper position the light on control panel will turn ON 3 If present test the jar switch by turning it left or right 4 Set vacuum regulator to position max 5 Close the VACUUM tube with a finger...

Page 7: ...ials only 1 Remove the lid of reusable canister 2 disassembly the cage of the overflow valve and the overflow valve 3 Clean components in hot water 60 70 C containing a detergent with a pH range betwe...

Page 8: ...ses and verifying smooth working operations 3 Check the mechanical integrity of the aluminium case the penetration of external material could damage the equipment or create danger to the operators 4 C...

Page 9: ...canister 20 kg 48 5 lbs approx max vacuum level 675 mmHg 90 kPa 0 9bar 5 CEI EN60601 1 Classification Insulation Class I device Applied part type B Device not suitable for use in presence of flammabl...

Page 10: ...pplicable Note 2 These guidelines may not apply to all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 12 SPARE PARTS Use only access...