TROUBLESHOOTING
Message
Level
Cause
Solution
Medium Temperature too low Warning
Process temperature of the medium
is too low.
Increase the process temperature
according to the specification (see
Medium Temperature too
high
Failure
Process temperature of the medium
is much too high.
Reduce the process temperature
according to the specification
Internal error
Failure
Internal device error. The sensor
goes into the safe state.
Please contact SICK.
Sensor is not taught
Warning
The sensor is in teach-in mode, but
there are no valid teach-in values.
The flow rate is 0%. The outputs are
switched accordingly.
Correct the minimum and maximum
value for teach-in.
Teach is not finished
Warning
Teach was started but did not com-
plete.
Perform the teach-in again.
8.2
Outputs
Error
Cause
Solution
The switching out-
put does not behave
as expected.
Configuration is incorrect.
Configure the switching output (see
An error has occurred; the sensor outputs are in
the safe state.
Eliminate the cause of the error.
Cable break.
Check the cable.