9. TECHNICAL DATA
Technical data
Electrical connection
Plug connector
M12 round connector x 1 (4-pin or 5-pin)
Short-circuit resistance
S+ / SP1 / SP2 to U–
Reverse polarity protection U+ to U–
Insulation voltage
DC 500 V
Over voltage protection
DC 40 V
Pin assignment
See type label
Materials
Parts in contact with
measuring material
< 10 bar (145 psi): 316L
≥ 10 bar (145 psi): 316L, PH steel
Parts not in contact with
measuring material
Housing
304
Keyboard
TPE-E
Display pane
PC
Display head
PC+ABS blend
Pressure transmission medium
Synthetic oil for all gauge pressure measuring ranges < 10 bar
(145 psi), all absolute pressure measuring ranges and flush-
mounted versions (< 16 bar (232 psi) with increased overload
protection).
Process connections
Standard
Threaded size
Overload limit
Seal
DIN 3852-E
G ¼ A
1,000 bar (14,500 psi) NBR
(FPM/FKM included)
EN 837
G ¼ female thread
1)
1,000 bar (14,500 psi)
–
ANSI/ASME B1.20.1
¼ NPT 1)
1,000 bar (14,500 psi)
–
G ½ B flush-mounted
1,000 bar (14,500 psi)
NBR (option: FPM/FKM)
Other process connections available with adapters.
1)
Optionally suited for oxygen, free of oil and grease.