Operation
MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG
77
Su
bj
ect
to
ch
an
ge
with
ou
t n
otic
e
5.7.11.7
Results
Specialist
: System edit: Results
In this menu, option results are calculated, i.e. the raw data are processed according to the
formula entered to obtain e.g. values of concentration, absorption, flow rate, volumes etc.
The drift values for QAL3 (
) can be processed and reset.
The “Results” menu is the basis for other menus which make use of the results (“R”)
calculated therein. In the latter menus, only a list containing the corresponding result
numbers is displayed from which you can make your selection.
Max. 64 results can be selected and activated.
The results are calculated every second.
As a default setting the results 1..24 correspond to the components 1..24. Linearization
and interference tables can be applied to these results.
Although the default setting can be modified, this option should only be used if the results
exceeding R24 do not suffice for the calculations.
If data are read-in via the serial interface
(Specialist: System edit: Data storage and send to
PC)
, then the data are fed to the results (secondary variable S1i with MCS protocol, Ri with
ModBus protocol). For a detailed description of the protocols please refer to
.
Key assignment:
Shift F3: Copy from other lines
Settings
Result
Mark (
) = result is calculated.
The mark cannot be selected, if there is an error in the formula.
Name
Name of component
Unit
Unit of component.
Formula
The calculations are entered in this column.
The length of the input buffer is 80 characters.
Arithmetic Operators
+
Addition
-
Subtraction
*
Multiplication
/
Division
^
Exponentiation
(,)
Parentheses, for bracketing terms
[,]
Brackets, for bracketing indices
,
Comma, to separate variables resp. indices
Primary variables
ST
(Data storage cycle / Sample Time)
AT (Actual
Time)
Secondary variables:
Ri
Results (i = 1..64)
Default settings are:
R1..R14: Sample components 1..14
R15: Flow meter
R16: O
2
sensor
R17..R24: Analog inputs 1..8
R25..R32: Temperatures
1
Upon receipt via Modbus protocol:
(
) “Results” Ri = “Formula” Ri)
AIi:
Analog input channel (i = 1..32)