Commissioning and operation
Su
bj
ec
t t
o
ch
an
ge
w
ith
ou
t n
ot
ice
4.3.6.4
Analog output
The analog signal can represent a variety of different output value sources (
):
Fig. 32
List of output value sources
To adapt the FLOWSIC30 to the different application conditions, the analog output has to
be configured. The adjustment of the analog output requires the change of various
parameters.
The output current Iout is calculated as follows:
Here, the following parameters have to be set (see also
):
Table 7
AO parameters
Fig. 33
AORangeLow and AORangeHigh
Further parameters such as damping, error value, gain and current offset can be set.
Fig. 34
Analog Output Configuration
Parameter
Description
Q
Output value source, as given in the drop-down list displayed in
above figure
AORangeLow
Lower range limit (Q value at 4 mA)
AORangeHigh
Upper range limit (Q value at 20 mA)
I
out
= 4 mA +
Q
-‐
AORangeLow
AORangeHigh
-‐
AORangeLow
·
16 mA