FLOWSIC100 Flare · Addendum to Operating Instructions · 8025365/V1-0/2020-05 · © SICK Engineering GmbH
Installation
4.8.1.2
Calculating the wetted part length for the probe version
Probe version: 1-path measurement
Fig. 30
Formulas for calculation of the wetted part length wL for the probe version
wLProbe NozzleLength
2
GasketThickness
LengthBallValve WallThickness
sin
---------------------------------------- 120 79
mm
sin
--------------------------- 206
+
4
mm FittingAdjust
–
+
+
+
+
=
Installation without ball valve (non-retractable)
Installation with ball valve (retractable)
wLProbe NozzleLength GasketThickness
WallThickness
sin
---------------------------------------- 120 79
mm
sin
--------------------------- 206
+
+
4
mm FittingAdjust
–
+
+
=
Example for 75°
wLProbe NozzleLength
2
GasketThickness
LengthBallValve WallThickness
75
sin
---------------------------------------- 120 79
mm
75
sin
--------------------------- 206
+
4
mm FittingAdjust
–
+
+
+
+
=
wLProbe NozzleLength
2
GasketThickness
LengthBallValve WallThickness
0 966
---------------------------------------- 330 45
mm
+
+
+
+
=
Example for 75°
wLProbe NozzleLength GasketThickness
WallThickness
75
sin
---------------------------------------- 120 79
mm
75
sin
--------------------------- 206
+
+
4
mm FittingAdjust
–
+
+
=
wLProbe NozzleLength GasketThickness
WallThickness
0 966
---------------------------------------- 330
+
45
mm
+
+
=
Simplified:
Simplified:
β [°] = Nominal nozzle installation angle
NozzleLength [mm] = Nozzle length D1 or D2; dimension on nozzle installation tool
GasketThickness [mm] = Dimension of gasket thickness
LengthBallValve [mm] = Length of ball valve, measure from gasket surface to gasket surface
WallThickness [mm] = Wall thickness of pipe
120,79 mm = Probe 1-path distance inner pipe wall to first probe A
82,189 mm = Probe 2-path distance inner pipe wall to first probe A
206,4 mm = Distance probe A to end of probe
330,45 mm = Probe 1-path wetted part length in pipe
290,48 mm = Probe 2-path wetted part length in pipe
0,966 = sin(75°) – Nominal nozzle installation angle
FittingAdjust [mm] = 1 mm Compensation when tightening the self-cutting ring fitting (self-cutting ring fitting displaces
the duct probe by 1 mm when tightened)