24
APPENDIX A. SPECIFICATIONS
TECHNICAL SPECIFICATIONS
AC test mode Test
frequency:
1 kHz, 10 kHz, 120Hz,100 Hz
Test frequency accuracy:
50 PPM (0.005%)
Test signal level:
0.25/0.5/1.0 +/- 5% Vrms Sine wave
Source impedance:
62.5
Ω
/1k
Ω
/16k
Ω
+/- 1%
TYPICAL OFFSET:
Resistance
≤
25 m
Ω
Capacitance
0.65 pF
Inductance
0.1 uH
Offset value should be subtracted from measurement result for
small value components (R < 10
Ω
, C < 100 pF, L < 10
μ
H).
MEASUREMENT RANGES
Parameter
Measurement Range Test frequency
Resistance
< 9.9 M
Ω
1 kHz
Capacitance
< 9999 pF
10 kHz
10000 pF to 1
μ
F
1 kHz
> 1 uF
100 Hz
Inductance
0.5
μ
H to 99
μ
H
10 kHz
100
μ
H 99 mH
1 kHz
> 100 mH
100 Hz