Response
time
0x0017
Write
Y
Read
Y
Description
Meter
response
or
data
sampling
time
Value
type
UINT
16
Notes
The
default
value
is
10
(10
msec).
Options
are
10,
20,
50,
100,
and
200
msec.
It
is
recommended
that
write
protection
should
be
applied
after
this
value
is
changed.
High
flow
alarm
0x0031
~
0x0032
Write
Y
Read
Y
Description
Set
the
high
flow
rate
alarm
limit
Value
type
UINT
16
Notes
The
set
flow
rate
value
=
[value
(0x0031)
*
65536+
value
(0x0032)]
/
1000
Low
flow
alarm
0x0033
~
0x0034
Write
Y
Read
Y
Description
Set
the
low
flow
rate
alarm
limit
Value
type
UNIT
16
Notes
The
set
flow
rate
value
=
[value
(0x0033)
*
65536
+
value
(0x0034)]
/
1000
Offset
reset
0x0027
Write
Y
Read
N
Description
Reset
or
calibrate
the
offset
Value
type
UINT
16,
fixed
value
0xAA55
Notes
Send
the
fixed
value
0xAA55
to
register
0x0027.
Please
make
sure
there
is
no
flow
before
executing
this
command.
Write
protection
0x0014
Write
Y
Read
N
Description
Write
protection
of
a
specific
value
Value
type
UINT
16,
a
fixed
value
of
oxAA55
Notes
This
function
disables
the
write
‐
protected
register
and
allows
you
to
modify
the
corresponding
register’s
value.
For
example,
to
change
the
gas
conversion
factor,
the
user
needs
to
send
0xAA55
to
the
register
0x0016,
and
then
the
write
function
will
be
enabled
(write
protection
is
disabled).
After
the
write
execution
is
completed,
the
firmware
will
automatically
re
‐
enable
the
write
protection.