Automatic well sounder “DUA-1”
15
control test. For the automatic valve the countdown to burst will start and
then the automatic valve will trigger.
For manual valve, the "Power" LED will start blinking once every
half second. After a second, generate an acoustic impulse by pressing the
manual valve knob of the product until it stops. The "Power" LED will
light up red.
Within 20 seconds, the product will record acoustic information from
the well, than determine the well fluid level and the "Power" LED will turn
green. The inscription "Survey is complete" will appear.
In order to import information to the "DB SIAM", enter the import
mode (tab "Devices") and press the "Import" button. If there is a large
amount of data, the "Instrument does not respond" message may appear.
To resume import from the interrupted place, press the "Import" button.
2.5.3. Automatic removal of level recovery or drop curve
Mount the product to the wellbore branch pipe.
For automatic removal of the recovery curve or level drop, set the
type of investigation LRC (level recovery curve) or LDC (level drop
curve). Set the acoustic pulse generation by solenoid valve. Set the period
of activation of the well sounder and the number of measurements in the
first interval. If one interval is not enough, set the necessary activation
periods and number of measurements in all necessary intervals. To start the
automatic mode, press the "Start" button on the "Start" tab.
In 5 seconds, on the "Information" tab, you will be able to observe
the current interval and the time before measurement. Then you can turn
off the product. The product will be switched on automatically with the
period specified in the first interval. After the number of measurements in
the first interval, the product will turn on with the period set in the second
interval, and so on until the fifth interval is completed or the next interval
is zero.
In automatic mode, the device records an echo together with a
symbolic report in a memory area designed to record 310 echoes. In this
case, the symbolic report is duplicated in the memory area designed to