6.4 Column Cleaning 

Problems in  peak shapes and elution time  changes or elevated column pressure etc. are often caused by 

insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 

problems may be resolved by cleaning the column.   

If a guard column is used with the analytical column, first remove the guard column and check the performance 

of the analytical column alone. If the problem is solved, most likely the cause is from the guard column. In this 

case, clean the guard column.   

If the problem is not solved by removing the guard column, clean both guard and analytical columns. Make sure 

to clean the guard and the analytical columns separately. When washing the column, let the washing solution 

flow from the column outlet go directly into the waste container and not let the solution go through the detector. 

 

<Cleaning Method> 

(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 

introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate. 

 

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction 

and use the flow rate at 0.35 mL/min 

 

Method 1: Adsorption of hydrophilic compounds   

Washing Solution 

Cleaning Time 

1  H

2

O   

25 min 

2  100 mM KOH or NaOH aq. solution 

100 min 

3  H

2

  25 min 

4  Eluent 

100 min 

 

 

Method 2: Adsorption of hydrophobic compounds 

Washing Solution 

Cleaning Time 

1  H

2

25 min 

2  H

2

O/CH

3

CN = 95/5 

20 min 

3  CH

3

CN 

100 min 

4  H

2

50 min 

5  Eluent 

100 min 

 

 

7. Column Storage

 

If preferred to store the column with the eluent used, remove the column from the system, securely tighten the 

end caps  and store the column in a refrigerator. Alternatively,  if preferred to store the column at a room 

temperature, replace the in-column solvent with the shipping solvent. Then, remove the column from the system, 

securely tighten the end caps and store the column at a location with stable temperature (a cool and dark space 

is recommended). Refer to section 6.3 Eluents for how to replace the eluent.   

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.   

· When replacing the in-column solvent with shipping solvent, detach the suppressor before 

introducing the shipping solvent to the column. 

   

 

· When reusing the column stored in a refrigerator, first wait for the column temperature to 

be ambient, and then remove the end caps and install the column on the system.     

 

 

 

 

Summary of Contents for IC SI-36 4D

Page 1: ...ckage adhesive label and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IC SI 36 4D is a suppressor type ion chromatography column for analysis of anions The column is used with hydroxide eluents such as potassium hydroxide and sodium hydroxide The column is suitable for the separations of standard anions including fl...

Page 2: ...e is less than 10 25 μL per column When analyzing low concentration samples 0 1 mg L increasing the injection volume to 100 200 μL may improve the detection 4 For the samples containing proteins remove the proteins prior to the sample injection The use of an ultrafiltration membrane is recommended 5 For the samples containing hydrophobic impurities remove the impurities prior to the sample injecti...

Page 3: ...ss than 0 35 mL min at the end of analysis Then turn off the column oven and let the column temperature return to room temperature before stopping the pump If the pump was stopped while the eluent inside the column was still hot as the eluent temperature decreases its volume also decreases This may result in creating an empty space in the column and deteriorates the column 6 3 Eluents 1 Use freshl...

Page 4: ...cleaning steps for adsorbing components For an efficient cleaning reverse the flow direction and use the flow rate at 0 35 mL min Method 1 Adsorption of hydrophilic compounds Washing Solution Cleaning Time 1 H2O 25 min 2 100 mM KOH or NaOH aq solution 100 min 3 H2O 25 min 4 Eluent 100 min Method 2 Adsorption of hydrophobic compounds Washing Solution Cleaning Time 1 H2O 25 min 2 H2O CH3CN 95 5 20 m...

Page 5: ...late Number N 2 Asymmetry Factor Fas 9 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support off...

Page 6: ... anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained...

Reviews: