Shodex IC SI-36 4D Operation Manual Download Page 3

ver.200603E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

6.2 Column Installation

 

 

(1) Connect the column to LC system by following the “flow direction arrow” (

) indicated on the adhesive label.   

 

(2) Make sure to insert the tubing all the way to the end fitting and secure it with the male nut. It is important that 

there is no extra space between the tubing and the column side of the end fitting. Presence of an extra space 

will let the sample to spread out and may result in wide peaks. 

 

 

 

 

(3) Set the initial flow rate at less than 0.35 mL/min and start the system. If using the column at an elevated 

temperature, keep a low flow rate until the temperature of the column reaches to the set temperature, and 

then gradually increase the flow rate to the desired.   

 

(4) When using the column for the first time, before connecting the column outflow to a suppressor, wash the 

column with the eluent at 0.35-mL/min flow rate for more than 30 minutes. This also applies to when reusing 

a column with which the eluent was replaced with shipping solvent.   

 

 

  Caution! 

 

 

· Verify that there is no solvent leak. It may cause electronic leakage, rust, and/or chemical 

injury. 

 

 

Attention!

   

· Make sure not to let air bubbles enter the column while installing the column. The air 

bubbles may damage the column.   

 

· When restarting the system after column installation or after holding the eluent flow, start 

the system at less than 0.35 mL/min. A rapid increase in pressure can damage the column.   

 

· If the column was used at an elevated temperature, lower the flow rate to less than 0.35 

mL/min  at the end of analysis. Then, turn off the column oven, and let the column 

temperature return to room temperature before stopping the pump. If the pump was 

stopped while the eluent inside the column was still hot, as the eluent temperature 

decreases, its volume also decreases. This may result in creating an empty space in the 

column and deteriorates the column. 

 

6.3 Eluents 

(1) Use freshly generated ultrapure water (

18.2 M

Ω

-cm

 ) to prepare the eluent.   

 

(2) The use of an eluent generation system is highly recommend for the preparation of potassium hydroxide or 

sodium hydroxide solution. If an eluent generation system is not available, dilute a reagent-special grade 

potassium hydroxide or sodium hydroxide solution with ultrapure water.     

 

(3) Degas the eluent fully to prevent the formation of air bubbles.   

 

(4) Presence of small debris or insoluble substances may result in deterioration of the column and/or they appear 

as noise on the chromatograms. Filter the eluent with a 0.45-

μ

m disposable filter to prevent the problems. 

 

(5) Start the system at less than 0.35 mL/min when replacing the solvent. 

 

 

 

Attention!

   

· Use HPLC grade organic solvent whenever possible. Solvents left in an opened bottle for 

a long time should not be used. The content may have been changed, absorbed moisture, 

or has been contaminated. 

 

· Always use freshly prepared solvents. Solvents stored for a long time may have changed 

their compositions and may influence elution patterns and/or damage the column. 

 

· If using a manually prepared eluent, keep the eluent in a solvent reservoir with a carbonate 

trap cartridge.   

 

     

Note  

   

· Use of on-line degasser is recommended. 

 

· It is recommended to set the pump limiter to avoid exceeding the maximum pressure.   

 

Summary of Contents for IC SI-36 4D

Page 1: ...ckage adhesive label and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IC SI 36 4D is a suppressor type ion chromatography column for analysis of anions The column is used with hydroxide eluents such as potassium hydroxide and sodium hydroxide The column is suitable for the separations of standard anions including fl...

Page 2: ...e is less than 10 25 μL per column When analyzing low concentration samples 0 1 mg L increasing the injection volume to 100 200 μL may improve the detection 4 For the samples containing proteins remove the proteins prior to the sample injection The use of an ultrafiltration membrane is recommended 5 For the samples containing hydrophobic impurities remove the impurities prior to the sample injecti...

Page 3: ...ss than 0 35 mL min at the end of analysis Then turn off the column oven and let the column temperature return to room temperature before stopping the pump If the pump was stopped while the eluent inside the column was still hot as the eluent temperature decreases its volume also decreases This may result in creating an empty space in the column and deteriorates the column 6 3 Eluents 1 Use freshl...

Page 4: ...cleaning steps for adsorbing components For an efficient cleaning reverse the flow direction and use the flow rate at 0 35 mL min Method 1 Adsorption of hydrophilic compounds Washing Solution Cleaning Time 1 H2O 25 min 2 100 mM KOH or NaOH aq solution 100 min 3 H2O 25 min 4 Eluent 100 min Method 2 Adsorption of hydrophobic compounds Washing Solution Cleaning Time 1 H2O 25 min 2 H2O CH3CN 95 5 20 m...

Page 5: ...late Number N 2 Asymmetry Factor Fas 9 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support off...

Page 6: ... anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained...

Reviews: