LBL 61876 rev. E, March 2018  

EN 

12 

 

6.3

 

Other Maintenance 

Shockwave Medical recommends that you contact your Shockwave Medical representative if you have any questions or concerns about maintenance. 

Shockwave Medical recommends replacement of IVL Connector Cables every three years to reduce the possibility of failure during patient use. In the event IVL Connector 

Cable connectors have become contaminated or the IVL Connector Cable malfunctions, remove this cable from use and contact your Shockwave Medical representative for a 

replacement. Additional IVL Connector Cables can be ordered separately. 

 

NOTE

 The IVL Connector Cable should not be disposed of in the normal waste stream; it should be sent to a separate collection facility for recovery and recycling. 

 

6.4

 

Product Useful Life 

The IVL Generator has been designed to provide a useful life of three years or more.  Life expectancy is based on actual usage. Periodic inspection per the maintenance 

schedule above is recommended by Shockwave Medical to assess continued use. 

 

7.

 

Troubleshooting 

If a problem is detected with the IVL System during use or testing, refer to the troubleshooting tips below. If the problem cannot be corrected, remove the equipment from 

service and contact your Shockwave Medical representative and or email [email protected].  

Technical Support: 

For Shockwave Medical Technical Support, contact your local Shockwave Medical representative or 

www.shockwavemedical

.com. 

 

WARNING – SHOCK HAZARD 

Do not attempt to service the system. It contains no operator serviceable components and dangerous high voltages may be present. No user modification or servicing of this 

equipment is allowed. If any part of this product appears damaged, remove from use and contact your Shockwave Medical representative for repair or replacement. 

 

Observation 

Possible Cause 

Corrective Action 

Unit will not turn on 

Battery requires charge 

Connect Charger Module to the IVL Generator and to AC 

Mains.   

Allow the IVL Generator to charge at least twelve (12) hours 

prior to use 

Battery will not charge  

(BATTERY CHARGING indicator is off) 

Disconnected cable 

Connect Charger Module to the IVL Generator and also to 

AC Mains.   

NOTE: two cable connections are required. 

Low BATTERY CAPACITY indicated when the 

battery symbol is Empty 

Battery requires charge 

Connect Charger Module to the IVL Generator and to AC 

Mains.   

Allow the IVL Generator to charge at least twelve hours 

prior to use 

System Error displayed and red light around 

POWER ON/OFF button. 

Built in tests have detected a malfunction within the IVL 

Generator 

Turn IVL Generator off, wait one second and turn the IVL 

Generator on again. 

NOTE: If SYS fault cannot be resolved, remove IVL 

Generator from service and contact your Shockwave Medical 

representative. 

Catheter Error Displayed

 

Catheter connection has not been made properly or has 

become disconnected. 

Make sure IVL Connector Cable is connected to IVL 

Generator   

Make sure valid IVL Catheter is connected to IVL Connector 

Cable 

IVL Catheter has reached its useful life or is defective  

Replace IVL Catheter 

IVL Connector Cable has reached its useful life 

Replace Connector Cable 

THERAPY STATUS indicator on IVL Generator front 

panel or on the IVL Connector Cable does not turn 

on 

Valid IVL Catheter is not connected 

Make sure IVL Connector Cable is connected to IVL 

Generator   

Make sure IVL Catheter is connected to IVL Connector Cable 

Replace IVL Catheter 

IVL Generator will not activate 

(THERAPY STATUS is off) 

Valid IVL Catheter is not connected 

See THERAPY STATUS indicator troubleshooting step above 

IVL Generator will not activate 

(THERAPY STATUS is yellow) 

THERAPY ON/OFF button is pressed 

Release THERAPY ON/OFF button, try again 

THERAPY STATUS changed from green to yellow 

IVL System has automatically paused therapy (see Section 

4.0 Step 10 

THERAPY STATUS should indicate green again, within the 

pause period specified in the IVL Catheter IFU 

THERAPY ON/OFF button on the IVL Connector 

Cable is pressed, but  IVL Generator does not 

deliver pulses 

IVL Generator cannot be activated  

(THERAPY STATUS indicator is off) 

If error message(s) appear, see troubleshooting guide 

above 

IVL Generator has not yet been activated (THERAPY STATUS 

indicator is yellow) 

Press THERAPY ON/OFF button once (THERAPY STATUS 

should turn green) 

IVL System has automatically paused therapy (THERAPY 

STATUS indicator is yellow, see Section 4.0 Step 9) 

THERAPY STATUS should indicate green again, within the 

pause period specified in the IVL Catheter IFU 

IVL Catheter or IVL Connector Cable is faulty 

(THERAPY STATUS indicator is green) 

Replace IVL Catheter 

 

Replace IVL Connector Cable 

 

 

 

 

Summary of Contents for IVL

Page 1: ...ion and Test 9 6 Maintenance 10 6 1 Daily Maintenance 10 6 1 1 Charging and Testing the Internal Battery 10 6 1 2 Testing the IVL Generator 11 6 1 3 Inspecting the IVL Generator 11 6 1 4 Cleaning the IVL Generator 11 6 2 Monthly Maintenance 11 6 3 Other Maintenance 12 6 4 Product Useful Life 12 7 Troubleshooting 12 8 Appendix A Electromagnetic Compatibility Guidance 13 9 Appendix B Symbols 14 10 A...

Page 2: ...result in serious personal injury or death CAUTION Hazards or unsafe practices that may result in minor personal injury product damage or property damage 2 2 Contraindications There are no specific contraindications for use of the IVL Generator and Connector Cable However users should read and understand the specific indications contraindications warnings and precautions included with the applicab...

Page 3: ...rature and humidity conditions for at least twenty four hours before use See Appendix C for specified operating conditions Operating the equipment outside of these environmental conditions may cause equipment malfunction or damage IMPROPER DEVICE PERFORMANCE HAZARDS Using other manufacturers cables catheters power adapters or batteries may cause the device to perform improperly and may invalidate ...

Page 4: ...ailable Counts down from available Pulse Count Per Catheter during treatment as each pulse is delivered Refer to applicable IVL Catheter IFU for the Max Pulse Count THERAPY STATUS Indicator On green when the device is ready to deliver therapy Flashes to indicate therapy is in process On yellow when therapy is paused or deactivated See Section 4 Steps 5 9 3 3 Front Panel Connectors FEATURE DESCRIPT...

Page 5: ...NECTOR end of the Connector Cable to the THERAPY CONNECTOR Orient the connector and gently push in The connector is magnetic and will engage as the magnet gets close Push gently to confirm connector is fully engaged Slide CONNECTOR DOOR to left to reveal the THERAPY CONNECTOR THERAPY CONNECTOR Step 5 Prepare IVL Catheter for Use Prepare the catheter for use following the instructions in the IVL Ca...

Page 6: ...nt as indicated in the IVL Catheter IFU CAUTION EQUIPMENT DAMAGE The IVL Generator delivers low energy short duration high voltage pulses to the IVL Catheter through the IVL Connector Cable The system is designed not to deliver pulses unless an IVL Catheter connector is mated with the IVL Connector Cable and IVL Generator It is important not to allow the contacts or internal surfaces of unmated co...

Page 7: ...unting the IVL Generator to an unstable IV pole could present a tip over hazard to staff or patients 5 1 IV Pole Mounting STEP 1 Locate and identify mounting hardware see image below Two identical sets of IV pole mounting hardware are provided One set is used to mount the IVL Generator and one set is used to mount the battery Charger Module IV Pole Mounting Hardware STEP 2 Determine the IV pole di...

Page 8: ...econd bracket so that it rests on top of the first bracket Orient the brackets so that they are facing opposite directions STEP 4 Mount the IVL Generator to the top bracket Align the holes in the mounting plate on the right side of the IVL Generator with the bracket mounting studs Push the IVL Generator towards the bracket to engage the mounting studs then push the IVL Generator down to lock in pl...

Page 9: ...orage so that the battery will not discharge to an unacceptably low voltage level which could ultimately result in battery damage 5 4 Environment Conditioning The IVL Generator is designed to be used indoors in a controlled environment Refer to Appendix C for specified operating conditions Allow the IVL Generator to be stored in the ambient conditions of the usage environment for at least 24 hours...

Page 10: ...IVL Generator Spilled liquids may cause the IVL Generator to perform inaccurately or malfunction 6 Maintenance This section describes the maintenance that clinical and or biomedical staff should be familiar with performing on a regular basis Recommendations for regular maintenance and maintenance intervals are noted below NOTE The battery is self discharging and requires periodic recharges even du...

Page 11: ... this button for three seconds is required Step 4 Confirm Output Test Result The output test requires approximately 15 seconds to complete During this time the THERAPY STATUS indicator will remain on green Upon successful completion of this test the IVL Generator will sound four beeps If no error message appears on the display then this test has passed successfully 6 1 3 Inspecting the IVL Generat...

Page 12: ...required Low BATTERY CAPACITY indicated when the battery symbol is Empty Battery requires charge Connect Charger Module to the IVL Generator and to AC Mains Allow the IVL Generator to charge at least twelve hours prior to use System Error displayed and red light around POWER ON OFF button Built in tests have detected a malfunction within the IVL Generator Turn IVL Generator off wait one second and...

Page 13: ... hospital environment If the user of the Generator Model 825DP requires continued operation during power mains interruptions it is recommended that the Generator Model 825DP be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typic...

Page 14: ...affected by absorption and reflection from structures objects and people Essential Performance Generator Model 825DP maintains safe and effective performance in the delivery of IVL therapy when operated in the electromagnetic environment specified in the table above 9 Appendix B Symbols The IVL Generator bears the following symbols Symbol Description Symbol Description Refer to instruction manual ...

Page 15: ...stance 10 mL saline from above Generator 100 mL saline from any angle Connector Cable distal end Weight 15 pounds 6 8 kg 10 2 Appendix C2 Performance Specifications This appendix contains the specifications and performance characteristics for the Shockwave Medical IVL Generator Model 825DP Specifications Performance Characteristics Battery Rechargeable Smart Lithium Ion Battery Pack 14 4V 6 6Ah Ch...

Page 16: ...PN 62393 Rev A Manufactured by Shockwave Medical Inc 48501 Warm Springs Blvd Suite 108 Fremont CA 94539 USA www shockwavemedical com patents MedPass SAS 95 bis Boulevard Pereire 75017 Paris France ...

Reviews: