LBL 61876 rev. E, March 2018  

EN 

 

 

 

 

4.

 

 Product Use and Therapy Delivery 

Before use read all sections of this operator’s manual and familiarize yourself with all controls, displays and connector features. Charge the IVL Generator before use (see 

Sections 5.2, 5.3). Also refer to the IFU provided with the IVL Catheter for additional information before use.  Not all clinical procedures will follow the sequence below. The 

following steps serve as a guide for use of the IVL Generator in clinical applications.  

Step  

Picture or additional info if applicable 

 

Step 1 – Turn the Generator On 

Momentarily depress the POWER ON/OFF button.  All indicators on the IVL Generator will come on briefly as a test.  The THERAPY STATUS 

indicator will come on yellow and green during this test.  The POWER STATUS indictor will continue to be on green. 

 

 

Step 2 – Confirm Battery Capacity 

With the Generator powered on, the BATTERY CAPACITY will be shown in the right side of the display.  The battery symbol should be at least 

partially filled as shown.   

 

If the battery symbol is empty, additional charging of the battery is recommended before use because there may not be an adequate charge to 

complete a patient treatment. 

 

 

 

Step 3 – Check Diagnostics 

Confirm that the display is as pictured with no error messages displayed.  If an error message is displayed, refer to Troubleshooting, Section 

7.0.   

 

Normal display with no errors shown to the right.

 

 

 

 

 

If a yellow light is displayed, refer to Troubleshooting, Section 7.0. 

 

If any error messages come on during use, refer to Troubleshooting, Section 7.0. 

 

Error Condition – Catheter error shown to the right. 

 

 

Step 4 – Connect IVL Connector Cable 

Disconnect the Charger Module from the CHARGER CONNECTOR if it is connected.  

  

Slide the CONNECTOR DOOR fully to the left, revealing the THERAPY CONNECTOR as shown. 

 

Connect the GENERATOR CONNECTOR end of the Connector Cable to the THERAPY CONNECTOR.  Orient the connector and gently push in.  

The connector is magnetic and will engage as the magnet gets close. Push gently to confirm connector is fully engaged. 

 

 

 

Slide CONNECTOR DOOR to left to reveal the 

THERAPY CONNECTOR 

 

 

THERAPY CONNECTOR: 

 

 

Step 5 – Prepare IVL Catheter for Use 

Prepare the catheter for use following the instructions in the IVL Catheter IFU.  Use a sterile sleeve to cover the distal end of the IVL Connector 

Cable.  Protect the connector from contamination by fluids. 

 

N/A 

 

Step 6 – Connect IVL Catheter 

Take care not to contaminate either connector end with fluids or other foreign matter during this procedure prior to mating. 

Connect the IVL Catheter to the CATHETER CONNECTOR end of the IVL Connector Cable using the same method outlined in step 4.   

 

NOTE: Ensure that the sterile sleeve also covers the CATHETER CONNECTOR.  

The THERAPY STATUS indicator on the IVL Generator front panel and the CATHETER CONNECTOR will be yellow, indicating that the IVL 

Generator is ready to activate. 

 

The IVL Catheter balloon dimensions will appear in the BALLOON SIZE display.  The total number of available pulses for the selected IVL 

catheter will appear in the PULSE COUNT field. 

 

 

 

Step 7 – Position the IVL Catheter 

Following conventional angioplasty catheter technique, introduce and position the IVL Catheter as desired. Use caution to prevent unintentional 

movement of the IVL Connector Cable and IVL Catheter during treatment.   

 

 

(Refer to the IVL Catheter IFU) 

 

Step 8 – Activate the IVL Generator 

Inflate the IVL Catheter and verify pressure per the instructions indicated in the IVL Catheter IFU. Press the THERAPY ON/OFF button once.  

The THERAPY STATUS indicator on the IVL Generator front panel and on the CATHETER CONNECTOR will now be green, indicating that the IVL 

Generator is now ready to deliver therapy. To deactivate the IVL Generator at any time, simply press the IVL Generator THERAPY ON/OFF 

button again and verify that the THERAPY STATUS indicator light is yellow.   

 

 

Summary of Contents for IVL

Page 1: ...ion and Test 9 6 Maintenance 10 6 1 Daily Maintenance 10 6 1 1 Charging and Testing the Internal Battery 10 6 1 2 Testing the IVL Generator 11 6 1 3 Inspecting the IVL Generator 11 6 1 4 Cleaning the IVL Generator 11 6 2 Monthly Maintenance 11 6 3 Other Maintenance 12 6 4 Product Useful Life 12 7 Troubleshooting 12 8 Appendix A Electromagnetic Compatibility Guidance 13 9 Appendix B Symbols 14 10 A...

Page 2: ...result in serious personal injury or death CAUTION Hazards or unsafe practices that may result in minor personal injury product damage or property damage 2 2 Contraindications There are no specific contraindications for use of the IVL Generator and Connector Cable However users should read and understand the specific indications contraindications warnings and precautions included with the applicab...

Page 3: ...rature and humidity conditions for at least twenty four hours before use See Appendix C for specified operating conditions Operating the equipment outside of these environmental conditions may cause equipment malfunction or damage IMPROPER DEVICE PERFORMANCE HAZARDS Using other manufacturers cables catheters power adapters or batteries may cause the device to perform improperly and may invalidate ...

Page 4: ...ailable Counts down from available Pulse Count Per Catheter during treatment as each pulse is delivered Refer to applicable IVL Catheter IFU for the Max Pulse Count THERAPY STATUS Indicator On green when the device is ready to deliver therapy Flashes to indicate therapy is in process On yellow when therapy is paused or deactivated See Section 4 Steps 5 9 3 3 Front Panel Connectors FEATURE DESCRIPT...

Page 5: ...NECTOR end of the Connector Cable to the THERAPY CONNECTOR Orient the connector and gently push in The connector is magnetic and will engage as the magnet gets close Push gently to confirm connector is fully engaged Slide CONNECTOR DOOR to left to reveal the THERAPY CONNECTOR THERAPY CONNECTOR Step 5 Prepare IVL Catheter for Use Prepare the catheter for use following the instructions in the IVL Ca...

Page 6: ...nt as indicated in the IVL Catheter IFU CAUTION EQUIPMENT DAMAGE The IVL Generator delivers low energy short duration high voltage pulses to the IVL Catheter through the IVL Connector Cable The system is designed not to deliver pulses unless an IVL Catheter connector is mated with the IVL Connector Cable and IVL Generator It is important not to allow the contacts or internal surfaces of unmated co...

Page 7: ...unting the IVL Generator to an unstable IV pole could present a tip over hazard to staff or patients 5 1 IV Pole Mounting STEP 1 Locate and identify mounting hardware see image below Two identical sets of IV pole mounting hardware are provided One set is used to mount the IVL Generator and one set is used to mount the battery Charger Module IV Pole Mounting Hardware STEP 2 Determine the IV pole di...

Page 8: ...econd bracket so that it rests on top of the first bracket Orient the brackets so that they are facing opposite directions STEP 4 Mount the IVL Generator to the top bracket Align the holes in the mounting plate on the right side of the IVL Generator with the bracket mounting studs Push the IVL Generator towards the bracket to engage the mounting studs then push the IVL Generator down to lock in pl...

Page 9: ...orage so that the battery will not discharge to an unacceptably low voltage level which could ultimately result in battery damage 5 4 Environment Conditioning The IVL Generator is designed to be used indoors in a controlled environment Refer to Appendix C for specified operating conditions Allow the IVL Generator to be stored in the ambient conditions of the usage environment for at least 24 hours...

Page 10: ...IVL Generator Spilled liquids may cause the IVL Generator to perform inaccurately or malfunction 6 Maintenance This section describes the maintenance that clinical and or biomedical staff should be familiar with performing on a regular basis Recommendations for regular maintenance and maintenance intervals are noted below NOTE The battery is self discharging and requires periodic recharges even du...

Page 11: ... this button for three seconds is required Step 4 Confirm Output Test Result The output test requires approximately 15 seconds to complete During this time the THERAPY STATUS indicator will remain on green Upon successful completion of this test the IVL Generator will sound four beeps If no error message appears on the display then this test has passed successfully 6 1 3 Inspecting the IVL Generat...

Page 12: ...required Low BATTERY CAPACITY indicated when the battery symbol is Empty Battery requires charge Connect Charger Module to the IVL Generator and to AC Mains Allow the IVL Generator to charge at least twelve hours prior to use System Error displayed and red light around POWER ON OFF button Built in tests have detected a malfunction within the IVL Generator Turn IVL Generator off wait one second and...

Page 13: ... hospital environment If the user of the Generator Model 825DP requires continued operation during power mains interruptions it is recommended that the Generator Model 825DP be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typic...

Page 14: ...affected by absorption and reflection from structures objects and people Essential Performance Generator Model 825DP maintains safe and effective performance in the delivery of IVL therapy when operated in the electromagnetic environment specified in the table above 9 Appendix B Symbols The IVL Generator bears the following symbols Symbol Description Symbol Description Refer to instruction manual ...

Page 15: ...stance 10 mL saline from above Generator 100 mL saline from any angle Connector Cable distal end Weight 15 pounds 6 8 kg 10 2 Appendix C2 Performance Specifications This appendix contains the specifications and performance characteristics for the Shockwave Medical IVL Generator Model 825DP Specifications Performance Characteristics Battery Rechargeable Smart Lithium Ion Battery Pack 14 4V 6 6Ah Ch...

Page 16: ...PN 62393 Rev A Manufactured by Shockwave Medical Inc 48501 Warm Springs Blvd Suite 108 Fremont CA 94539 USA www shockwavemedical com patents MedPass SAS 95 bis Boulevard Pereire 75017 Paris France ...

Reviews: