Indication when input value is -200.0 (-2000) or less
When the range has a decimal point: For the indication of -200.0 or less, the input value
and the minus (-) sign are indicated alternately.
When DC voltage or current input is selected, the indication of -2000 or less is the
same as the above.
(E.g.) Indication of -200.0
IN
IN
•
Indication when input value is 10000 or more
When DC voltage or current input is selected: For the indication of 10000 or more, the
lower 4 digits of input value are flashing.
(E.g.) Indication of 10020
IN
•
Underrange, Overrange and Sensor Burnout indication
Underrange : If input value becomes -10% of the input span or less, "
" flashes on
the Input display.
Overrange : If input value becomes 110% of the input span or more, "
" flashes
on the Input display.
•
Indication time setting
If indication time is set, the displays will go off after the indication time has elapsed.
(Only the power indicator remains lit.)
If power is turned ON again, or if any of the keys
,
,
or the SUB-MODE Key
is pressed while displays are unlit, the displays will light again.
Flashes.