5 User information setting
FZ-Ex series operetion manual
-
42
-
5-8-2
Actual value setting method
1
Select a setting menu.
Select the discrimination value setting of the
comparator function.
(“2-7-2 Operation of the setting menu, setting of various
functions”)
2
Select the reference value setting. (In the case
of the relative value discrimination)
Push the [Direction] key.
Select “37.TS.”
Push the [Direction] key.
Select “371.TG.”
Push the [Enter] key.
3
Set a reference value.
Push the [Function F] key.
Place an object to be weighed that is
equivalent to the reference value.
Push the [Enter] key.
The reference value is stored.
4
Select the upper limit value setting.
Push the [Direction] key.
Select “37.TS.”
Push the [Direction] key.
Select “372.HI.”
Push the [Enter] key.
5
Set an upper limit value.
Push the [Function F] key.
Place an object to be weighed that is
equivalent to the upper limit value.
Push the [Enter] key.
The reference value is stored.
6
Select the lower limit value setting.
Push the [Direction] key.
Select “37.TS.”
Push the [Direction] key.
Select “373.LO.”
Push the [Enter] key.