5.1 Start-Up Inspection and Periodic Inspection
5-5
IRAffinity-1
The intensity specifications are given below.
1) There is an absorbance peak of water at 400 cm
−
1
, so correct intensity of the power spectrum
is not calculated. Thus the intensity of 403 cm
−
1
is used for evaluation of the power spectrum.
Fig. 5.4 Power Spectrum (Example)
If the intensity does not meet the specifications, or the energy at high wavenumbers (to 4600 cm
−
1
) is extremely
low, run [Auto adjustment] on the [Instrument] Menu, and reanalyze the spectrum. If the power spectrum does
not improve after the "auto adjustment", the system may be damaged; contact your SHIMADZU representative.
If the intensity remains zero, the light source may be disconnected or damaged; light source replacement or
repair should be performed by a SHIMADZU service engineer.
Wavenumber (cm
−
1
):
Specification
4600:
10% or more of maximum
4000:
25% or more of maximum
3000:
50% or more of maximum
Wavenumber indicating the maximum intensity
(2500 - 1800 cm
−
1
):
50 or more
700:
10% or more of maximum
500:
2% or more of maximum
403
1)
:
0.5% or more of maximum
351:
0.01% or more of maximum
Summary of Contents for IRAffinity-1
Page 2: ...This page is intentionally left blank...
Page 14: ...xii IRAffinity 1 Symbol Definition Current AC Symbols Found on the IRAffinity 1...
Page 39: ...2 1 IRAffinity 1 Chapter 2 Specifications CONTENTS 2 1 IRAffinity 1 Specifications 2 2...
Page 60: ...Chapter 3 Installation 3 18 This page is intentionally left blank IRAffinity 1...
Page 131: ...3 IRAffinity 1 Z ZnSe 7 2 7 4...
Page 132: ...4 This page is intentionally left blank IRAffinity 1...
Page 134: ...This page is intentionally left blank IRPrestige 21...