47
9.2.2.
Service Mode Errors
When entering service mode, detailed error codes are displayed for the errors generated in user
mode, as shown in the table below. In this state, press the
[→] key to display a history of the
most recent errors, up to a maximum of five errors.
When more than five errors are displayed, press the
[→] key to return to the display of the most
recent errors.
If entering service mode when no errors have been generated, "NO ERR" is displayed, as
shown in the table below.
User Mode Error
Name
Service Mode Error Name
Display
Cause
No Errors
-
No errors occurred during user mode.
System error
(Hardware error)
Weight loader system error
(Err0)
Weight loader system for the ATX
Series did not function correctly due to
malfunction, etc.
Temperature sensor error (Err1)
Abnormal temperature data due to
temperature sensor malfunction
EEPROM data error (Err5)
Abnormal mass calculation due to
EEPROM data abnormality
Sensitivity
calibration error
Stabilization error (CAL E1)
Measurement value at zero point or
during weight measurement is
unstable.
Zero point error (CAL E2)
Drift in zero point from last sensitivity
calibration is large. During first
sensitivity calibration after energizing,
the zero point change exceeds 20 % of
the weight capacity, or a later zero point
change exceeds 2 % of the weight
capacity.
Reference weight error (CAL E3)
Difference between reference weight
value input during PCAL and measured
weight span value exceeds 2 % of
reference weight value.
Span error (CAL E4)
Difference between the set reference
weight value and measured weight
span value exceeds 2 % of the
reference weight mass.