17
mode.
·
Connect the test leads across with the object being
measured.
·
Read the value from display.
Warning
a) Do not apply any voltage to the input when the Meter is
working in resistance measuring mode.
b) If resistance of the circuit or resistor be measured is
open or higher than the range be selected the display
will show “OL”.
c) For low resistance measurement, it maybe make some
additional error due to the resistance of the test leads, at
this case, as a tip you can short the test leads firstly and
record down the resistance, then subtract it from the
measuring value to obtain the more accurate result. If
the short resistance of the test leads is up to 0.5
Ω
, it is
need to double check if the test leads are loosened or
other troubles exist.
d) It is normal for taking a couple of seconds to get the
steady reading when you do the resistance
measurement greater than 1M
Ω
.
e) When the measurement has been completed,
disconnect the connection between the testing leads and
the circuit under test and remove testing leads from the
input terminals.