13
The accuracy of SpO
2
and PR measurement still
meet the precision described above when the
modulation amplitude is as low as 0.6%.
P. Ambient Light Interference:
The difference between the SpO
2
value measured
in the condition of indoor natural light and that of
darkroom is less than ±1%.
Q.
Dimensions:
56 mm (L) × 34 mm (W) × 30 mm (H)
Net Weight:
approx. 60g (including batteries)
R. Classification
The type of protection against electric shock:
Internally powered equipment.
The degree of protection against electric shock:
Type BF applied parts.
The degree of protection against harmful solid
foreign objects and ingress of liquid:
The equipment is IP22 with protection against
harmful solid foreign objects and ingress of liquid.
Electro-Magnetic Compatibility:
Group I, Class B
5 Packing List