After turn on this function, the peristaltic pump will measure the volume
automatically, when the volume reach set up volume, the pump will stop
working automatically. The flow rate can be changed during the pump working.

Click the

Fixed Volume

button, set

ON

to turn on this function. Click

Set Volume

,

to input volume, the unit can be mL or L, range is 0.01mL to 9999L. The
prompting frame display the needed time to finish the volume with set up flow
rate. The maximum time is 9999min, when more than 9999min, the system will
warn.

11

Dispensing Interface:

Dispensing Interface as below:

After turn on this function, the peristaltic pump will transfer fixed volume within
set up time. This function can be used for one time liquid dispensing. Click the

Summary of Contents for V6-12L Series

Page 1: ...Shenchen Peristaltic Pump Manual of V6 12L Series...

Page 2: ...in time Connect directly the power line to the wall socket and avoid using the extense electric line If the power line or plug had wear and other damage please hold the plug to unplug it not hold the...

Page 3: ...rd Instruction 2 V6 12L Series Operation Interface Structure 3 V6 12L Series Operation Interface Instruction 3 V6 12L Series Technical Specification 16 V6 12L Series Function and Features 16 External...

Page 4: ...splays in the same screen Intelligent calibration and online micro adjusting function three measurement mode fixed volume measurement fixed time and volume timer start and stop It can load YZ35 pump h...

Page 5: ...highest speed This button can be used for wash tube or fast filling liquid CW CCW Button Press this button the motor will change running direction When the pump working with fixed volume measurement...

Page 6: ...Series Operation Interface Instruction 1 Boot Interface After power on the system enter the welcome interface choose system language Chinese English If do not choose it will enter the English main in...

Page 7: ...er When these two functions turn off display the parameter is all 0 Set Parameter Display Display the fixed volume measurement fixed time and volume state information the model of pump head and tube s...

Page 8: ...Display input units when input flow rate or volume Input Digital Area Numeric keypad Unit Clr Button When input flow rate or volume this button is unit switch you can choose different unit When it is...

Page 9: ...flow rate mode or rotating speed mode button to choose the working mode When you choose the flow rate mode the flow rate is adjustable the speed will change with the flow rate When you choose the rota...

Page 10: ...ontrol to enter Extrocontrol Settings interface A According to the external port speed analog input signal to select signal 0 5V 0 10V 4 20mA analog signal three kinds of optional analog signal voltag...

Page 11: ...ic pump address No range 1 32 select communication enable is ON Then this pump can be communication with master receiving master signal NOTE Peristaltic pump only under communication control when in t...

Page 12: ...No back to the calibration interface C After click the Start button during the pump running you can click the Stop button to stop the test D After finish the tests the actual volume area display the a...

Page 13: ...face as below Click the System Setting button from the main interface click Date and Time button enter date and time setting interface The date and time will display on the top right corner of main in...

Page 14: ...set ON to turn on this function Click Set Volume to input volume the unit can be mL or L range is 0 01mL to 9999L The prompting frame display the needed time to finish the volume with set up flow rat...

Page 15: ...is 0 360 Click Run time input working time The prompting frame display the flow rate and motor speed with the set up parameters 12 Timer Start And Stop Interface Timer Start And Stop Interface as bel...

Page 16: ...rocess 1 Flow Rate Transferring Function Note Flow rate calibration process please refer the flow rate calibration interface instruction 2 Fixed Volume Measurement Function Note Flow rate calibration...

Page 17: ...Note Flow rate calibration process please refer the flow rate calibration interface instruction 4 Timer Start And Stop Function Under the flow rate transferring mode set the pump start at 8 30 a m fro...

Page 18: ...ristaltic Pump V6 12L Series 15 Click Start Time set the start time is 8 30 a m turn the button to ON Click Custom come out the repeat date window as below Timer stop setting process is same with the...

Page 19: ...color touch screen control dynamic display working status the flow volume and motor speed display in the same screen Intelligent calibration function it can calibrate the flow rate and dispensing vol...

Page 20: ...industrial production Timer start and stop function Can set the pump start and stop time freely reach automation control Load shedding memory function store the running parameters in time safe and re...

Page 21: ...d turn on the Ext Speed Control the motor speed from 0 rpm to 600rpm through analog signal 0 10V 0V to 10V voltage signal input terminal 0 5V 0V to 5V voltage signal input terminal 4 20mA 4mA to 20mA...

Page 22: ...irection signal input Pulse mode the direction change once when receive one pulse signal rising edge effectively Level mode when high level the pump clockwise running when low level the pump countercl...

Page 23: ...p running Level mode R S 1 short connect with GD1 the pump running when disconnect the pump stop running This terminal can connect with passive switch and foot pedal In the external control setting in...

Page 24: ...ignal terminal RXD Peristaltic pump sending master receive signal terminal RS232 Communication Interface Connection Diagram as below RS485 Communication Interface Choose RS485 in the communication set...

Page 25: ...Shenchen Peristaltic Pump V6 12L Series 22 Product Dimension mm...

Page 26: ...therwise it can quicken the tubing wearing reduce the useful life of tubing and lead the rollers to damage in earlier Pump head can not resist super corrosive liquid Please pay attention to it when it...

Page 27: ...n the warranty the products are damaged because of users wrong operation or other human damages our company do not responsibility for warranty Beyond the warranty we only charge the cost of maintenanc...

Page 28: ...ng Shenchen Precision Pump Co Ltd Address No 267 KeYuan Street MinYing Technology Park Baoding China Tel 86 312 2055881 2031189 Fax 86 312 2055371 Website www good pump com Email easypump2 gmail com e...

Reviews: