Shenchen Peristaltic Pump
V6-12L Series
- 9 -
time and volume. Others display transferring mode.
If fixed time and volume turn on, the target volume and running time is set up
parameter, unable to amend. Other modes, the running time is 60s, you can
click the run time button to amend the running time.
Before the pump working, need to calibrate the flow rate to ensure the
transferring or dispensing accuracy
Process as below:
A
、
Confirm the running time, if fixed time and volume function, the running time is
set up time, unable to change.
B
、
Click
Start
button to test, countdown display the run time, it will stop
automatically, and display numerical keyboard, input the actual volume, then it
will ask whether continue test ( suggest more than 3 times), choose
Yes
, the
pump will test again, choose
No
, back to the calibration interface.
C
、
After click the
Start
button, during the pump running, you can click the
Stop
button to stop the test.
D
、
After finish the tests, the actual volume area display the average data, click the
CAL
button, the calibration is finished. Now the request flow rate or volume is
close to theoretical data.
E
、
If request higher accuracy, you can click
Add
and
Dec
button to micro adjust the
flow rate, to reach high accuracy transferring and dispensing.
F
、
Click
Cancel
button, cancel the test data, the actual volume return to 0.
Online Micro Adjust Volume Process:
Flow Rate Transferring Mode:
If the actual flow rate is big or small than the set
up flow rate, you can micro adjust the flow rate online without affect the product
line.
Fixed Time and Volume Mode
: If the dispensing volume is big or small than the
set up volume, you can micro adjust the volume online, no need to stop the pump.
Fixed Volume Measurement Mode:
Do not support online micro adjust function.