20.0

20.0

EZ1

EZ2

1 2 3

C

ACTUAL PROCESS 

TEMPERATURE DISPLAY

PROCESS SETPOINT 

DISPLAY

INDICATES OUTPUT 1 ACTIVE

HEATING ELEMENT ON

INDICATES OUTPUT 2 ACTIVE

LIGHTS ON

INDICATES OUTPUT 3 ACTIVE

LIGHT CALIBRATION OFFSET ON

RAMPING SYMBOL

INDICATES PROGRAM 

RUNNING WHEN ACTIVE

ADVANCE KEY

ALLOWS TO SCROLL

THROUGH PARAMETER LIST

DOWN ARROW KEY

ALLOWS TO LOWER SETPOINT 

OR CHANGE PARAMETERS

UP ARROW KEY

ALLOWS TO RAISE SETPOINT

OR CHANGE PARAMETERS

INFANITE KEY

ALLOWS TO BACK UP

ONE LEVEL OR RETURN 

TO HOME PAGE

EASY ZONE KEY 1

STARTS AND STOPS 

PROGRAM

CONTROLLER KEYS AND DISPLAYS

 

 

5.5 

Set Temperature Control and AM/PM Light Control:

  

The control comes from the 

factory with a 12-hour day and a 12 hour night cycle already programmed into the contol.  To 
run  the  program,  choose  the  Set  Point  that  you  desire  by  using  the  UP  or  DOWN  Arrow 
Keys.    (Bottom  Display  Set  Point  and  Top  Display  Process).    Once  Set  Point  has  been 
entered, push the EZ1 Button and the program will start.  The Program is set up to run the 
day cycle first.  To stop the program, push the EZ1 Button again.  (Note:  When stopping the 
program, the control will remain in the condition that it is previously in.)

 

 

To restart the program, push the EZ1 Button again.  (Note:  The 12 Hour Day Cycle always 
starts  first  and  begins  when  the  program  is  started.)   Day cycle is running with Indicator 2 
and 3-Ramp Symbol is illuminated. 

5.6 

Controller  Programming  Overview:   

The  controller  is  capable  of  programming  four 

(4) different files with ten (10) different steps in each file or one (1) file with forty (40) different 
steps.  The 12-hour day/12-hour night cycle using one Set Point uses one (1) file with four 
(4) steps.  The second step run the 12-hour day cycle.  After the 12-hour timer counts down, 
it automatically switches to Step 3 which runs the night cycle.  After the 12-hour timer counts 
down, it automatically switches to Step 4.  Step 4 tells the program to return to Step 2 and 
the process starts all over.  Step 1 is an unused step reserved for Specific Time activation.  
Before a Specific Time activation time is entered, the Real Time Clock must be set for the 
real time.  Next are instructions on “How To Set The Real Time Clock” and “How To Enter A 
Program at Specific 

Time.” 

Summary of Contents for RF02033111

Page 1: ...LOW TEMPERATURE DIURNAL ILLUMINATION INCUBATOR MODEL RF2015 Serial Number RF02033111 INSTALLATION AND OPERATION MANUAL REV 09 10 4861428...

Page 2: ...ROUBLESHOOTING SECTION 8 0 PARTS LIST UNIT SPECIFICATIONS WIRE DIAGRAMS This unit is a general purpose Diurnal Growth Chamber for professional industrial or educational use where the preparation or te...

Page 3: ...r will arrange for official inspection to substantiate your claim 1 3 Return Shipment Save the shipping crate until you are sure all is well If for any reason you must return the unit first contact yo...

Page 4: ...r circulation around the unit allow a minimum of 20cm between incubator and any walls or partitions which might obstruct free air flow 2 3 Lifting Handling These units are heavy and care should be tak...

Page 5: ...when cleaning and assure all volatile or flammable cleaners are evaporated and dry before reattaching the unit to the power supply...

Page 6: ...tes that you should consult your manual for further description and discussion of a control or user item 2 Indicates Temperature 3 Indicates Overtemperature 4 C Indicates Degrees Centigrade 5 Indicate...

Page 7: ...ain Controllers set point If temperature rises to the Over Temperature set point this thermostat takes control of the heating element and allows continued use of the incubator until the problem can be...

Page 8: ...ower switch to the ON position and turn each Overtemperature Thermostat to its maximum position clockwise using a coin or flat edged tool 5 2 Place a certified reference thermometer not supplied in th...

Page 9: ...he day cycle first To stop the program push the EZ1 Button again Note When stopping the program the control will remain in the condition that it is previously in To restart the program push the EZ1 Bu...

Page 10: ...EXT BOTTOM DISPLAY SCREEN PUSH UP OR DOWN ARROWS TO SELECT HOUR OF DAY IN UPPER DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN PUSH UP OR DOWN ARROWS TO SELECT MINUTES OF HOUR IN UPPER DISPLAY PUSH...

Page 11: ...HANGE UPPER DISPLAY TO HOUR OF DAY PUSH ONCE FOR NEXT SCREEN USE UP OR DOWN ARROWS TO CHANGE UPPER DISPLAY TO MINUTE OF HOUR HOUR OF DAY MINUTE OF HOUR PUSH ONCE FOR NEXT SCREEN USE UP OR DOWN ARROWS...

Page 12: ...H ONCE FOR NEXT BOTTOM PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN TO SELECT S tyP IN TOP DISPLAY USE UP OR DOWN ARROWS TO SELECT 12 IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN PUSH ONCE FOR...

Page 13: ...IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT oFF IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT 4 IN TOP DISPLAY PU...

Page 14: ...rent Set Point Temperatures This program uses one 1 file and six 6 different Steps Step 1 Is an unused Step and is reserved for starting the program at Specific Time Step 2 Parameters are for entering...

Page 15: ...S TO SELECT DESIRED SET POINT IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN TO SELECT 0 IN TOP DISPLAY USE UP OR DOWN ARROWS TO SELECT 0 IN TOP DISPL...

Page 16: ...OTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT on IN TOP DISPLAY USE UP OR DOWN ARROWS PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT 0 IN TOP DISPLAY hoUr hoUr 11 Min...

Page 17: ...PLAY SCREEN Ent 1 USE UP OR DOWN ARROWS TO SELECT oFF IN TOP DISPLAY PUSH ONCE TO RETURN PREVIOUS SCREEN S tyP USE UP OR DOWN ARROWS TO SELECT SoAh IN TOP DISPLAY S tyP SoAh PUSH ONCE FOR NEXT BOTTOM...

Page 18: ...FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT 2 IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT 0 IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DIS...

Page 19: ...N TO NORMAL DISPLAY A1 oPEr OR UNTIL SCREEN APPEARS AFTER THE OFFSET HAS BEEN ENTERED ALLOW UNIT TO STABALIZE RECHECK TEMPERATURE AGAIN AND IF CALIBRATION IS WITHIN EXCEPTABLE LIMITS THE OFFSET VALUE...

Page 20: ...OR UNTIL SCREEN APPEARS nonE Usr S Usr S SET 1 PUSH ONCE PUSH TWICE 20 0 20 0 TO RETURN TO NORMAL DISPLAY SELECTING SET 1 UNDER Usr S SAVES THE OFFSET CALIBRATION FOR THE NIGHT CYCLE BEFORE CALIBRATIO...

Page 21: ...in the chamber will spike and the Main temperature Controller will cycle off shutting down the heating element When the compressor is activated the temperature will stabilize at set point 6 2 Cleanin...

Page 22: ...connection check shadow box for loose connections Unit will not heat over a temperature that is below set point 1 confirm that fan is moving and that amperage and voltage match data plate check fan mo...

Page 23: ...ometer don t match 1 calibration error see section 5 9 2 controller failure evaluate if pilot light is operating correctly 3 allow at least two hours to stabilize 4 see if reference thermometer is cer...

Page 24: ...not miss aligned 2 Steady internal clicking may be broken spring or valve call Customer Service MECHANICAL Motor doesn t move 1 if shaft spins freely check connections to motor and check voltage to m...

Page 25: ...ce 2 check fuse circuit breaker on unit or in wall 3 see if unit is on e g fan or heater and just controller is off 4 check all wiring connections especially around the on off switch Contamination in...

Page 26: ...t 4660505 4660505 High Limit Control 1750862 1750862 Interior Outlet 6100525 6100526 Interior Outlet European NA 6100531 Low Limit Control 1750538 1750538 Main Temperature Control 1750855 1750855 Pilo...

Page 27: ...ons Shipping Weight 400 lbs Net Weight 280 lbs Exterior LxDxH in 34 5 x 32 x 77 Interior LxDxH in 27 x 23 5 x 57 Capacity 20 3 Cubic Ft Capacity 257 Bottles Temperature Range 10 to 40 C Temperature Un...

Page 28: ...RED22G BLK22G SSRELAY 4 3 RTD AC 98 K2 W1 K3 99 L2 S1 R1 B5 D6 D5 L3 J3 DC Y1 NEUTRAL TERMINAL BLOCK BLU BLU RED RED YEL POWER SWITCH 9851057 dlr rev 6 22 09 POWERRELAY 115V 7030511 230V 7030520 7850...

Reviews: