30 |
P a g e
USER MAINTENANCE
Continued next page
Temperature Calibration Continued
3
Place the display in its temperature calibration mode.
a.
Press and hold both the
UP
and
DOWN
temperature arrow buttons simultaneously for
approximately 5 seconds.
b.
Release the buttons when the temperature display
shows the letters “C O”. The display will begin
flashing the
current temperature display value
.
Note:
If an arrow key is not pressed for five seconds, the
display will cease flashing, and store the last displayed number
as the new current chamber temperature value.
4
Use the
Up
or
Down
arrows to adjust the current
display temperature value until it matches the
reference device temperature reading.
Reference Device
5
After matching the display to the reference device, wait 5
seconds.
•
The temperature display will cease flashing and
store the corrected chamber display value.
•
The incubator will now begin heating or
passively cooling in order to reach the set point
with the corrected display value.
Cooling to Set Point
6
After
the incubator has achieved the corrected
temperature, allow the chamber to sit at least one 1 hour
undisturbed to stabilize.
•
Failure to wait until the incubator is fully
stabilized will result in an inaccurate reading.
Set Point Achieved
Summary of Contents for SMI11
Page 1: ...SMI INCUBATORS 100 120 Voltage Installation Operation Manual SMI2 SMI6 SMI7 SMI11 SMI12...
Page 2: ...2 P a g e Pictured on cover SMI6 SMI2 SMI7 SMI11 SMI12 2 Stacked SMI6s...
Page 14: ...14 P a g e INSTALLATION...
Page 16: ...16 P a g e GRAPHIC SYMBOLS...
Page 18: ...18 P a g e CONTROL PANEL OVERVIEW...
Page 32: ...32 P a g e USER MAINTENANCE...