29
|
P a g e
Begin Calibraiton
USER MAINTENANCE
5) Temperature Stabilization
The incubator air temperature must stabilize in order to perform an accurate calibration.
•
Allow the incubator to operate undisturbed with the doors shut for
at least 8 hours
when
first putting the unit into operation in a new environment.
•
To be considered stabilized, the incubator chamber must operate at your calibration
temperature for
at least 1 hour with no fluctuations of ±0.1°C or greater.
Continued next page
Suggested Temperature Calibration
1
Once the incubator temperature has stabilized, compare the
reference device and incubator temperature display readings.
•
If the readings are the same, or the difference between the
two falls within the acceptable range of your protocol, the
display is accurately showing the incubator air
temperature.
The Temperature Calibration procedure is
now complete
.
- Or -
•
If a difference falls outside of your protocol range, advance
to step 2.
Reference Device
SET TEMPERATURE
2
A display calibration adjustment must be entered to match the
display to the reference device. See next step.
Reference Device
Required temperature stabalization period operating undisturbed with the incubator door(s) closed.
Start
Fluctuations
(Exaggerated)
Summary of Contents for SMI11
Page 1: ...SMI INCUBATORS 100 120 Voltage Installation Operation Manual SMI2 SMI6 SMI7 SMI11 SMI12...
Page 2: ...2 P a g e Pictured on cover SMI6 SMI2 SMI7 SMI11 SMI12 2 Stacked SMI6s...
Page 14: ...14 P a g e INSTALLATION...
Page 16: ...16 P a g e GRAPHIC SYMBOLS...
Page 18: ...18 P a g e CONTROL PANEL OVERVIEW...
Page 32: ...32 P a g e USER MAINTENANCE...