32 |
P a g e
OPERATION
Continued from the previous page.
The incubator is now ready for use.
•
into the incubator. Please see page 40.
8.
Run the unit for at least
12 hours
(for example, overnight) with
chamber doors closed prior to:
•
Setting the Over Temperature Limit (next step).
•
Loading samples.
This helps ensure a stable temperature environment.
9.
Set the Over Temperature Limit
procedure on
page 39.
•
The incubator must be heated
and stable
at your
application temperature prior to performing this
procedure.