INSTALLATION
P
OWER
S
OURCE
R
EQUIREMENTS
When selecting a location for the unit, verify each of the following requirements is satisfied.
Power Source
: The power source must match the voltage and amperage requirements listed on the
unit data plate. This unit is intended for
220 – 240 volt,
50/60 Hz
applications at
3.5 amps.
•
The wall power source must be protective earth grounded
.
•
The unit may be damaged if the supplied voltage varies by more than 10% from the data
plate rating.
•
Use a separate circuit to prevent loss of the unit due to overloading or circuit failure.
•
The recommended wall circuit breaker for this unit is 20 amps.
•
The wall power source must conform to all national and local electrical codes.
Power Cord:
The unit must be positioned so that all end-users can quickly unplug the cord in the
event of an emergency.
•
The unit is provided with a
230-volt, 10 amp, 2.5 meters (8ft 2in
),
Euro CEE 7/7
power cord.
Always use this cord or an identical replacement.
Fuses:
The unit ships with a fuse installed in the power cord inlet and a second fuse installed in an
adjacent fuse holder.
•
Both fuses must be installed and intact for the unit to operate.
•
Always find and fix the cause of a blown fuse prior to putting the unit back into operation.
•
Fuse type:
o
250V T6.3 Amp, 5x20mm
Standard
CEE7/7 wall
socket