OPERATION
S
ET THE
T
EMPERATURE
S
ET
P
OINT
End of Procedure
1
. Set OTL control to its maximum setting, if not already set to max.
Turning the OTL all the way to the right (clockwise) prevents the heating cutoff
system from interfering with this procedure.
2.
Navigate to the Temperature Set Point Adjustment mode
Push either button
Set Point Adjustment Mode
Note
: The display will automatically exit the adjustment mode after 5 seconds of
inactivity on the arrow keys, saving the last shown set point value.
3.
Set the Temperature Set Point
New Set Point
Adjust
Note
: Adjust the set point to its lowest setting (OFF)
to turn off heating in the chamber.
4.
Wait for 5 seconds after entering the Set Point
Set Temperature
⁰C
The display will stop flashing. The set point is now saved in the controller.
See the
on page 39 for how to set the OTL system once the
incubation chamber has stabilized at your application temperature set point and
after you have performed any display verifications or calibrations.
HEATING ACTIVATED
Initial Set Point
Set Temperature ⁰C
Set Temperature ⁰C