     SADP_Operating_ Instructions_0114 

      

 

 

 

12 Problem 

Guide 

 
 
 

     

PROBLEM 

   

     

CAUSE

   

 

 

SOLUTION

 

 
 

ERRATIC  READINGS 

Static on Indicator 

Moisten indicator face with 50/50 mixture of 

 

 

 

 

 

 

 

detergent & water, or use proprietary anti-static 

 

       cleaner. 

DO 

NOT 

POLISH 

INDICATOR 

FACE. 

 
 
FULL SCALE READING 

Wet gas 

 

Stop gas supply & switch analyser off. 

 
 

   Short 

Circuit 

Sensor 

Disconnect plug from sensor & if the indicator 

 

 

 

 

Cable or Plug   

still reads over full scale deflection.  Cure the 

 

 

 

 

 

 

 

short circuit in the cable or plug or replace. 

 
 
 
 

 

 

 

Short Circuit Sensor 

Disconnect the plug from the sensor & note the 

 

 

 

 

 

 

 

indicator falls to 0 deflection. Apply approx 24V 

 

 

 

 

 

 

 

DC to  sensor terminals MOMENTARILY with  

 

 

 

 

 

 

 

the sensor in a dry condition.  Polarity is not 

 

 

 

 

 

 

 

important, but the contact must be brief or the  

 

       sensor 

may 

be 

damaged.  If this fails to clear the 

 

       fault 

the 

sensor 

must 

be 

replaced. 

 
 
ZERO  READING   

Open Circuit on Cable  Disconnect plug from sensor and connect centre 

 

 

 

 

 

 

 

pin of plug to the outer connection.  No reading 

 

       will 

be 

obtained. 

 

Repair 

cable. 

 
 

 

 

 

Open Circuit on Sensor  Test as for open circuit on cable, but note that  

 

 

 

 

 

 

 

reading over FSD obtained.  Check sensor 

 

       connection 

or 

replace 

sensor. 

 
 

   Autocal 

Wrongly 

Set. 

Readjust 

to correct setting (i.e. carry out the 

 

       Autocal 

procedure 

as 

on 

page 

4). 

 
 

 

 

 

Instrument  

 

Disconnect coaxial cable from instrument and 

 

 

 

 

Unserviceable   

connect short circuit across terminals and note  

 

 

 

 

 

 

 

no reading even with Autocal control turned up. 

 

       Return 

instrument 

for 

service. 

 

 

 

 

 

 

 

 

 

 

WET  READING   

Leak in System or 

Cure  leak, or replace unsuitable pipe with 

 

 

 

 

use of Unsuitable Pipe  copper or stainless steel.  Flexible connections 

should be made with PTFE pipe. NEVER use 
rubber or plastic pipe. 

 
 

   Comparison 

with 

 Ensure 

that samples are taken from same sample 

 

   Secondary 

Standard 

 

point. 

 

Re-check Autocal setting.  If problems still 

 

   Instrument 

or 

  occurs 

change 

sensor. 

 

 

 

 

Calibration Gas  

 

 

Summary of Contents for SADP

Page 1: ...SADP_Operating_ Instructions_0114 Model SADP Automatic Dewpoint Meter Operating Instructions...

Page 2: ...ing g_ Instructi Sh ons_0114 haw Moist Len S B t 44 f 44 e mail www s ture Meter Shaw Build Bolton Lane Bradford BD2 1AF England 0 1274 73 0 1274 37 shawmete shawmeter rs UK Ltd ding e 33582 70151 ers...

Page 3: ...ident please notify the carrier and the Shaw Moisture Meters representative from where this unit was purchased You should have received if ordered 1 Model SADP Instrument 1 Carry Case 1 Screwdriver Al...

Page 4: ...dex 4 4 0 Purging the Sample Connection 5 4 1 Sample Connections 5 5 0 Controls 5 6 0 Normal Operation 6 7 0 Calibration Checking 7 8 0 The Sensor 8 8 1 Gaining Access to the Sensor 8 9 0 Additional I...

Page 5: ...ternal power source The instrument comes supplied with 6 c size cells In normal operation these will last for over 12 months The unique measuring head is designed to keep the sensor dry when the instr...

Page 6: ...which effectively means that the instruments may be used in any environment except a deep mine The ATEX certification is Ex II 1 G Ex ia IIC T6 Ga This means Ex European certification mark for electri...

Page 7: ...is directly affected by changes in partial pressure of water vapour and these changes that are proportional to the dew frost point temperature are displayed on the instrument indicator 3 1 Gases to A...

Page 8: ...the air gas sample contains heavy hydrocarbon condensate the filter must be of the coalescing type with a drain The filter unit should be positioned as close to the sample point as practical 4 Pressu...

Page 9: ...re from the sample pipe work and components Check that no liquid or particulate contamination which could damage the sensor is passing through the sample pipe Directing the sample onto a piece of whit...

Page 10: ...the instrument head and slowly raise the head by hand 6 The instrument reading will move up the scale wetter and stabilise When there is no further change in reading note the result If the reading mo...

Page 11: ...d position 3 After about 1 minute not critical but not more than a few minutes check the instrument reading On SADP models it should be at the Automatic Calibration line It is not critical within 1 or...

Page 12: ...the dielectric value of the sensor which is then measured by the analyser A dynamic equilibrium will exist between the water vapour outside the sensor and the condensed water within the pores This equ...

Page 13: ...insert the head into the seal and when it meets the seal push firmly till it passes the seal 2 Unscrew the plain brass base of the inner section discard the old desiccant and replace with new 3 Check...

Page 14: ...lculator are printed on it 9 6 Response Time The response time of any water vapour pressure detector will naturally be very much quicker from dry to wet than from wet to dry This is the reason that th...

Page 15: ...st be cooled in order that it should be saturated with water vapour i e 100 relative humidity For practical reasons it is referred to water above 0 deg C and ice below 0 deq C PARTS PER MILLION BY VOL...

Page 16: ...s extremely low and so there is no hazard 6 Adjust the Automatic Calibration control to give a reading of full scale i e on the Automatic Calibration line 7 Remove the short circuit and the reading sh...

Page 17: ...ust be replaced ZERO READING Open Circuit on Cable Disconnect plug from sensor and connect centre pin of plug to the outer connection No reading will be obtained Repair cable Open Circuit on Sensor Te...

Page 18: ...If this does not cure problem return for calibration Comparison with Refer to Section 7 0 of manual re check Autocal Secondary Standard If this does not cure problem return for calibration Instrument...

Page 19: ...SADP_Operating_ Instructions_0114 13 Hygrometric Equivalents...

Page 20: ...meter showing Dewpoint and ppm with amplified range of 0 10ppm on TR versions SADP D 2cm 3 1 2 Digit LCD with battery check symbol Sensor Connection Internal Power Supply 9V DC 6 x VARTA 2014 R14 cell...

Page 21: ...SADP_Operating_ Instructions_0114...

Reviews: