OVER TEMP dial is placed inside the control panel.
OVER TEMP should be set at least 30℃ higher than the target set temp.
If the TEMP CONTROLLER set value is 120℃, OVER TEMP Dial must set over 150℃.
(Default value is 200 ℃)
<Position of OVER TEMP Dial Inside Control Panel>
AT
ON
OUT
MODE
TEMP & TIME CONTROL SYSTEM
AC
DOOR
P
A
1cm
And put water around 1cm in the tank.
Always check the level of the water before use.
Fill the water inside the steam bottle around 2cm.
2cm
<AC-60M/80M/100M>
<AC-128M>
INSTALLATION
9
<AC-60M/80M/100M>
<AC-60M/80M/100M>
⤵
⤵
Lock
Unlock
⤵
⤵
Lock
Unlock
*