(4) Start the unit
(5) Stop the unit
(6) Auto Turning(AT)
When TIME mode ends or operator stops it manually by pressing the unit will stop
running and in tank pressure will be normalized.
If the press is not normalized, please open pressure normalizing valve.
Make sure that you should press always to run the unit.
Lamp A and P's status is nothing to do with this product so you do not
need to pay attention for A/P Lamp status in this product.
Be careful about the hot steam.
Press for 3 sec and the unit will start running with lamp ON.
SETTING AND OPERATION
11
AT is recommended under 2 cases.
case1. When the change of current temp is fluctuating more than ±0.1℃.
case2. When the unit can not reach at the set temp.
You can enter / quit AT mode by pressing AT button for 5 sec.
AT will be finished automatically when it is done.
AT Operation
Display
② AUTO TUNING(AT) S/W
-Press and hold for 5 sec to
operate/quit
① Press
⤵
⤵