AUTOCLAVE
SH-AC-60M/80M/100M/128M
Version : 2016 Rev.0
Page 1: ...AUTOCLAVE SH AC 60M 80M 100M 128M Version 2016 Rev 0...
Page 2: ...Use the unit only in the way described in this manual Failure to follow the instruction in this manual may cause wrong operation 1 General Description 3 2 Graphic Simbols 4 4 The Outward 6 3 Key Featu...
Page 3: ...hoosing our quality product Our stands behind our products and wants to let you know we are here if you need us Before you use the unit read this entire manual carefully to understand how to install o...
Page 4: ...E OPERATING THE UNIT This symbol indicates items and procedures that are STRICTY PROHIBITED with regards to the unit Important operating and or maintenance instruction Read the accompanying text caref...
Page 5: ...when timer ends Perfect sterilization Auto back up AT ON OUT MODE TEMP TIME CONTROL SYSTEM AC DOOR P A Over temperature protection function Over current protection function Optimized pressure guage f...
Page 6: ...ler ELB breaker Electronic power cable Pressure gauge Drain valve Steam bottle Over pressure relief hole Pressure normalizing valve Locker nut PRESSURE GAUGE0 2 0 25 0 3 0 1 0 15 0 05 0 MPa CL 1 5 K A...
Page 7: ...s delivered place the package on a flat location Lift the box as shown in the picture Unscrew the bottom of the box Unwrap the bag or Place the equipment on your desired location DOOR Lift the remaini...
Page 8: ...normalizing valve is on close position before use It will cause pressure failure if the valve is not closed Serial label shows the electric power specification of the unit Connect power and turn on t...
Page 9: ...ust set over 150 Default value is 200 Position of OVER TEMP Dial Inside Control Panel AT ON OUT MODE TEMP TIME CONTROL SYSTEM AC DOOR P A 1cm And put water around 1cm in the tank Always check the leve...
Page 10: ...n Step1 From TEMP MODE press the MODE button for 1 sec to move to TIME MODE Step2 Set your desired time Step3 Press MODE button for 3 sec to get out of mode setting or press MODE button for 1sec to mo...
Page 11: ...h this product so you do not need to pay attention for A P Lamp status in this product Be careful about the hot steam Press for 3 sec and the unit will start running with lamp ON SETTING AND OPERATION...
Page 12: ...on one more time You can return to the original display by clicking AT button one more time 0 100 100 0 60 60 0 30 30 Output of heater Output of heater Output of heater SETTING AND OPERATION 12 7 Chec...
Page 13: ...L 100 L Dimension Internal 350 x630H mm 400 x650H mm 450 x650H mm 13 78 x24 80H inch 15 75 x25 59H inch 17 72 x25 59H inch Dimension External W D H 550x680x1060 mm 600x730x1060 mm 650x780x1080 mm 21...
Page 14: ...pe 0 to 0 3 MPa Sensor PT 100 Capa 128 L Dimension Internal W D H 500 x650H mm 19 69 x25 59H inch Dimension External W D H 700x830x1350 mm 27 56x32 68x53 15 inch Net weight 132 9 Kg Material Internal...
Page 15: ...emperature does not rise Press for 3 sec Press AT button for 5 sec Check OVER TEMP set value Pressure does not rise Press for 3 sec Press AT button for 5 sec Check OVER TEMP set value Check Pressure N...
Page 16: ......