8
2.0
The Control of the
RT 03
2.1
Switching on the device
The start up procedure of the
RT 03
is the following:
"
connect the piezoelectric microphone with the microphone
socket (pos. 5),
"
screw the appropriate adapter (square adapter or
horseshoe magnet) to the microphone,
"
attach the microphone to the fitting or the pipe,
"
connect the antenna (pos. 9) and
"
switch-on the device (pos. 1).
#
##
##
Do not switch-on the device without antenna !
#
##
##
Immediately after the switching on of the device, the automatic
level adjustment is carried out for about 6 secounds. During this
time you should avoid to make noise
(taps, voices or the like). The number
of bar segments on the bar chart shows
the number of operating hours still
available.
The
RT 03
now is transmitting and
permanently signals the relation
between the actual and the adjusted
level.
By means of the headphones (pos. 8), you can listen to the noise
coming from the fitting. Now, all adjustments of the transmitter
having been made, you can begin with the correlation.