51
7 Appendix
AL2
Type:
Main alarm
Adjustable:
Yes
Latching:
Yes
Trigger:
Alarm threshold AL2 exceeded
Indicator:
– Audible signal
– Visual signal
–
AL2
message on display
Acknowledge-
ment:
– Possible for audible signal when alarm threshold
AL2 is exceeded
– Possible overall after level falls below alarm thresh-
old AL2
Reset:
– By acknowledgement after level falls below alarm
threshold AL2
– By switching device off
Exception for
oxygen
: alarm AL2 is non-latching.
AL3
Type:
End of measuring range
Adjustable:
No
Latching:
Yes
Trigger:
Alarm threshold AL3 exceeded
Indicator:
– Audible signal
– Visual signal
–
AL3
message on display
–
Reading flashes
Acknowledge-
ment:
– Possible after level falls below alarm threshold AL2
Reset:
– By acknowledgement after level falls below alarm
threshold AL2
– By switching device off