51
Sample of Inspection Sheet (Methane-Devices)
Sample of Inspection Sheet (Methane-Devices)
Sample of Inspection Sheet (Methane-Devices)
Sample of Inspection Sheet (Methane-Devices)
Sample of Inspection Sheet (Methane-Devices)
TEST REPORT
EX-TEC® SR5, SR4, SR2, VARIOTEC® 8
Calibration:
methane CH4
Serial Number (e.g.: 041 11 0001)
23.10.1998
1.0
Device status
1.1
- status correct (e.g.: Y / N)
1.2
- fine-dust filters correct (e.g.: Y/ N)
1.3
- remaining operating hours (e.g.: 5 h)
2.0
Pump test
2.1
- low pressure > 150 mbar
2.2
- volume flow > 50 l/h
3.0
PPM measuring range
3.1
zero point
- fresh air reading
3.2
test gas 10 PPM CH4
- display > 5 PPM
3.3
test gas 100 PPM CH4
- display 70 ... 140 PPM
3.4
test gas 1.000 PPM CH4
- display 800 ... 1.200 PPM
3.5
test gas 1,00 VOL.% CH4
- display 0,80 ... 1,20 VOL.%
4.0
%LEL measuring range
4.1
zero point
- display -3 ... +3 %LEL or
- display -0,15 ... +0,15 VOL.%
4.2
test gas 50 %LEL = 2,20 VOL.% CH4
- display 45 ... 55 %LEL or
- display 2,00 ... 2,40 VOL.%
4.3
optical alarm (e.g.: Y/ N)
4.4
audible alarm (e.g.: Y / N)
5.0
VOL.% measuring range
5.1
zero point
- display -2 ... +2 VOL.%
5.2
test gas 100 VOL.% CH4 and/or Erdgas
- display 98 ... 102 VOL.%
6.0
Observations
- housing broken
- adjustment, repair
- factory inspection
- or the like
7.0
Test
- day
- month
- year
- signature